Deltabar M PMD55; (Technische Information) - PEI-FRANCE.com
Deltabar M PMD55; (Technische Information) - PEI-FRANCE.com
Deltabar M PMD55; (Technische Information) - PEI-FRANCE.com
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Deltabar</strong> M <strong>PMD55</strong><br />
Cellule de<br />
mesure<br />
10 mbar (0,15 psi)<br />
30 mbar (0,45 psi)<br />
100 mbar (1,5 psi)<br />
Temps mort<br />
[ms]<br />
Constante de temps T63 (= t 2 )<br />
[ms]<br />
500 mbar (7,5 psi)<br />
1 bar (15 psi)<br />
60 90<br />
3 bar (45 psi)<br />
16 bar (240 psi)<br />
40 bar (600 psi)<br />
Comportement dynamique :<br />
HART<br />
Temps mort, constante de temps<br />
Cellule de<br />
mesure<br />
Temps mort<br />
[ms]<br />
Constante de temps T63 (= t 2 )<br />
[ms]<br />
10 mbar (0,15 psi)<br />
30 mbar (0,45 psi)<br />
100 mbar (1,5 psi)<br />
500 mbar (7,5 psi)<br />
1 bar (15 psi)<br />
370 90<br />
3 bar (45 psi)<br />
16 bar (240 psi)<br />
40 bar (600 psi)<br />
Cycle de lecture<br />
Commande HART : en moyenne 2 à 3 par seconde<br />
Le <strong>Deltabar</strong> M maitrise la fonctionnalité BURST MODE, pour une transmission cyclique des valeurs via le<br />
protocole de <strong>com</strong>munication HART.<br />
Temps de cycle (temps de mise à jour)<br />
En moyenne 310...520 ms.<br />
Temps de réponse<br />
250 ms<br />
Amortissement<br />
Un amortissement agit sur toutes les sorties (signal de sortie, affichage digital).<br />
• Via l'affichage local, un terminal portable ou un PC avec logiciel d'exploitation progressivement de 0...999 s<br />
• Via micro-<strong>com</strong>mutateurs sur l'électronique, position "on" (= valeur réglée) et "off" (= amortissement inactif)<br />
• Réglage usine : 2 s<br />
Endress+Hauser 9