TaBle 2. Alarm messages (continued)<br />
When you see it means: do this:<br />
this message:<br />
32<br />
O 2 SENSOR<br />
P CIRC<br />
(High circuit<br />
pressure)<br />
O 2 %<br />
(High delivered<br />
O 2 %)<br />
V TE<br />
(High exhaled<br />
tidal volume)<br />
V .<br />
E TOT<br />
(High exhaled<br />
total minute<br />
volume)<br />
Background checks<br />
have detected a<br />
problem with the O 2<br />
sensor.<br />
The measured airway<br />
pressure is equal to or<br />
greater than the set<br />
limit. Reduced tidal<br />
volume likely.<br />
The O 2 % measured<br />
during any phase of a<br />
breath cycle is 7% (12%<br />
during the first hour<br />
of operation) or more<br />
above the O 2 % setting<br />
for at least 30 seconds.<br />
(These percentages<br />
increase by 5% for four<br />
minutes following a<br />
decrease in the O 2 %<br />
setting.)<br />
The patient’s exhaled<br />
tidal volume for any<br />
breath is equal to or<br />
greater than the set<br />
limit.<br />
The patient’s expiratory<br />
minute volume is equal<br />
to or greater than the<br />
set limit.<br />
O 2 sensor is out of<br />
calibration or has<br />
failed. Press 100%<br />
O 2 CAL; replace or<br />
disable the sensor.<br />
Check patient,<br />
patient circuit and<br />
endotracheal tube.<br />
Check patient, air<br />
and oxygen supplies,<br />
oxygen analyzer<br />
and ventilator.<br />
Check patient and<br />
settings. Consider<br />
whether the patient’s<br />
compliance or<br />
resistance has<br />
changed.<br />
Check patient<br />
and settings.<br />
<strong>Puritan</strong> Bennett <strong>840</strong> <strong>Ventilator</strong> <strong>Pocket</strong> <strong>Guide</strong>