Quick Installation Guide<br />

Step 3 continued...<br />

14<br />

Type n approvals : non-sparking / energy limited power supply<br />

<strong>Rosemount</strong> <strong>5300</strong> <strong>Series</strong><br />

Transmitter<br />

Field Communicator<br />

RS-485 with Modbus communication<br />

HART -<br />

HART +<br />

Fieldbus<br />

Modem<br />

Power<br />

Supply<br />

March 2013<br />

The <strong>5300</strong> <strong>Series</strong> RS-485 with Modbus communication transmitter version<br />

operates using a power supply ranging from 8 - 30 Vdc (max. rating). See the<br />

<strong>Rosemount</strong> <strong>5300</strong>/5400 <strong>Series</strong> with HART to Modbus Converter Manual<br />

supplement (Document No. 00809-0500-4530) for details.<br />

Power Consumption:<br />

< 0.5 W (with HART address = 1)<br />

< 1.2 W (incl. four HART slaves)<br />

HART to M odbus Converter<br />

MB<br />

MA<br />

MODBUS<br />

(RS-485) -<br />

- POWER +<br />

+<br />

Ambients > 60 ºC<br />

HART<br />

Use wirin g rated<br />

for min 90 ºC<br />

nverter<br />

120 <br />

MB<br />

MA<br />

MODBUS<br />

(RS-485)<br />

NOTE:<br />

<strong>Rosemount</strong> <strong>5300</strong> <strong>Series</strong> Transmitters with Flameproof/Explosion-proof output have a built-in<br />

barrier; no external barrier needed.<br />

-<br />

If it is the last<br />

transmitter on the<br />

bus, connect the<br />

120termination<br />

resistor.<br />

Power<br />

Supply<br />

A<br />

120 RS-485 Bus 120<br />

B<br />

PC