➔ Laboratory automation: see Chapter 9 ➔ Samplers: see Chapter 10 ➔ Process measurement technology: see Chapter 11-20<br />

Reliable and robust turbidimetry<br />

in the field<br />

➔ Standard-compliant technology<br />

➔ Precision due to RATIO optical system<br />

➔ Simple to operate<br />

➔ Reliable and robust<br />

➔ Fully equipped for on-site<br />

operation<br />

ISO with white light or infrared<br />

In order to be able to compare turbidity<br />

values reliably, traceability to international<br />

standards is necessary. The<br />

2100P ISO was designed to conform<br />

with EN ISO 7027 (860 nm IR LED).<br />

The 2100P model conforms completely<br />

to the guidelines of the USEPA 180.1.<br />

2100P portable turbidimeters<br />

Rough terrain and difficult<br />

samples—no problem<br />

The RATIO optical system guarantees<br />

precise measurement results right<br />

across the measuring range – even<br />

when samples are turbid or ultraclear.<br />

The non-adjustable optical system is<br />

enclosed in high-impact ABS plastic,<br />

which protects the instruments<br />

perfectly, even in the most difficult<br />

surroundings.<br />

MODEL 2100P 2100P ISO<br />

Art. no. 4650000 4474002<br />

Light source White light Infrared (860 nm)<br />

Standard USEPA 180.1 EN ISO 7027<br />

RATIO-Mode 90/180° 90/180°<br />

Measuring range 0.01 – 1,000 NTU 0.01 – 1,000 FNU<br />

Resolution 0.01 in the range from 0.01– 9.99<br />

0.1 in the range from 10.0 – 99.9<br />

1 in the range from 100 –1,000<br />

Accuracy ± 2 % of the reading plus scattered light<br />

Scattered light < 0.02 FNU/NTU<br />

Reproducibility ±1 % or 0.01 FNU/NTU<br />

Signal averaging Can be switched on or off<br />

Calibration 4-point calibration, automatic<br />

Response time 6 sec<br />

Display LCD<br />

Sample volume At least 15 ml<br />

Dimensions 22.2 x 9.5 x 7.9 cm (L x W x D)<br />

Weight 0.45 kg (2.7 kg in case with accessories)<br />

Operating conditions 0 – 50 °C<br />

www.greyhoundchrom.com<br />

Fully equipped in the field<br />

The easy to use 2100P and 2100P ISO<br />

turbidimeters can be powered by mains<br />

or batteries, and are factory calibrated.<br />

In addition, the practical carrying case<br />

holds everything needed for transporting,<br />

calibrating and measuring in the field.<br />

More information at www.hach-lange.co.uk,<br />

www.hach-lange.com, keyword “2100 Turbidity”,<br />

with free downloads of brochure for 2100P ISO<br />

(DOC062.52.00620) and User Manuals for 2100P<br />

(DOC022.52.00641) and 2100P ISO<br />

(DOC022.52.00630)<br />