Kobold Messring GmbH Chapter 3 Analogue Operation<br />

CHAPTER 3: ANALOGUE OPERATION<br />

Your <strong>DMS</strong> instrument may be operated in three different<br />

ways:<br />

THREE CONTROL OPTIONS<br />

A. Analogue Input/Output Operation (This Chapter): Using<br />

analogue input/output signals at the 15-pin mini-D connector.<br />

B. <strong>Digital</strong> Operation with Pilot Module (Chapter 4):<br />

Using the optional Pilot Module.<br />

C. <strong>Digital</strong> Operation with RS-232 <strong>and</strong> <strong>DMS</strong> Software (Chapter 5):<br />

Using the RS-232 link, the supplied <strong>DMS</strong> Software package<br />

<strong>and</strong> a PC-style computer running the Windows operating system.<br />

This chapter will discuss the first of these—Analogue Operation. Please<br />

see subsequent chapters <strong>for</strong> other options.<br />

Regardless of control options, the st<strong>and</strong>ard output <strong>for</strong> all <strong>DMS</strong> instruments<br />

are two linear analogue output signals corresponding to 0% to 100% of<br />

the mass flow full-scale range. Please note that one of these output signals<br />

is always 4-20 mA. The other is user-configurable to 0-5 VDC, 0-10 VDC or<br />

1-5 VDC.<br />

For mass flow controllers, one input signal of 4-20 mA, 0-5 VDC, 0-10<br />

VDC or 1-5 VDC (selectable by user) may be used to set the gas mass flow<br />

rate to any desired value within the range of the device. This input signal<br />

must be a direct linear representation of 0% to 100% of the desired gas mass<br />

flow full-scale value. For the location of these signals on the HD DB-15<br />

connector, refer to Figure 2-4: Wiring Definitions <strong>for</strong> Optional Communication<br />

Cable in Chapter .<br />

2<br />

<strong>DMS</strong> 3-1