GB<br />
Alarm Symbol Ca<strong>use</strong> Remedy<br />
Battery<br />
pre-alarm<br />
in battery<br />
operation mode<br />
Battery alarm in<br />
battery mode<br />
Alarm cannot be<br />
silenced whilst in<br />
battery mode<br />
Reminder<br />
Signal repeated<br />
every minute<br />
Target volume<br />
reached<br />
Plug symbol not visible in display<br />
Minimum battery voltage is<br />
not available. Pre-alarm<br />
sounds at least 30 minutes<br />
be<strong>for</strong>e battery alarm and<br />
pump stand by.<br />
Plug symbol not visible in display<br />
Minimum battery voltage is<br />
not available. Alarm comes<br />
on 10 minutes be<strong>for</strong>e<br />
batteries are completely<br />
discharged.<br />
<strong>Pump</strong> is switched on but not<br />
operated <strong>for</strong> a minute.<br />
Target volume reached (feed<br />
container symbol in display is<br />
flashing).<br />
Incorrect target volume<br />
inadvertently set.<br />
Alarms<br />
Connect pump to an<br />
electrical outlet via the pump<br />
holder. Recharge battery and<br />
operation of the pump can<br />
resume.<br />
Connect pump to mains via<br />
the pump holder. Recharge<br />
battery and operation of the<br />
pump can resume.<br />
Start pump or switch it off.<br />
End administration or<br />
continue feeding. If no target<br />
volume is desired,<br />
deactivate target volume<br />
function by pressing the<br />
bottom two arrow keys twice.<br />
Correct target volume.<br />