Guidance and manufacturer’s declaration<br />
ELECTROMAGNETIC EMISSIONS – TABLE 201<br />
Technical in<strong>for</strong>mation<br />
The <strong>APPLIX</strong> pump is intended <strong>for</strong> <strong>use</strong> in the electromagnetic environment specified below. The<br />
<strong>use</strong>r of the <strong>APPLIX</strong> pump make sure it is <strong>use</strong>d in such an environment.<br />
Emissions test<br />
RF emissions<br />
CISPR 11<br />
RF emissions<br />
CISPR 11<br />
Harmonic<br />
emissions<br />
IEC 61000-3-2<br />
Voltage<br />
fluctuations<br />
Flicker emissions<br />
IEC 61000-3-3<br />
Compliance<br />
obtained<br />
by the device<br />
Group 1<br />
Class B<br />
Complies<br />
Class A<br />
Does not apply<br />
Electromagnetic environment - guidance<br />
The <strong>APPLIX</strong> pump <strong>use</strong>s RF energy only <strong>for</strong> its internal function. There<strong>for</strong>e, its<br />
RF emissions are very low and are not likely to ca<strong>use</strong> any interference in nearby<br />
electronic equipment.<br />
The <strong>APPLIX</strong> pump is suitable <strong>for</strong> <strong>use</strong> in all establishments, including domestic<br />
and hospital establishments and those directly connected to the public lowvoltage<br />
power supply network that supplies buildings <strong>use</strong>d <strong>for</strong> domestic<br />
purposes.<br />