16 of 46 Setting up the <strong>HHPC</strong>-2 - <strong>HHPC</strong>-2 <strong>Handheld</strong> <strong>Particle</strong> <strong>Counter</strong><br />

<strong>Operator</strong> <strong>Manual</strong><br />

b ) Selecting Count Mode<br />

1) On the Main Screen, press the Mode key once.<br />

The bottom line displays the current mode.<br />

2) To review the modes available (listed in Table 2-1), press either the Start/Stop or the<br />

Mode key.<br />

Table 2-1 : Count Modes<br />

Fig 2-4 : Select Mode Screen<br />

Mode Description Sample Unit<br />

Concentration This mode is useful for taking a quick snapshot of<br />

airborne particulate samples, especially in areas where<br />

particulate levels are unknown and may exceed the<br />

operating levels of the counter. The <strong>HHPC</strong>-2 estimates<br />

count per cubic foot or liter depending on the volume<br />

selected. The unit updates the results by calculating<br />

completed sample counts and volume compared to the<br />

balance of volume remaining.<br />

Rate Counting is continuous. The displayed counts are a ten<br />

sample sliding average.<br />

Totalize The number of counts are displayed as they accumulate<br />

until the end of the sample. When the sample is<br />

completed, the record is stored and the value is displayed<br />

on the screen until the next sample is started.<br />

Volume size<br />

Volume unit<br />

Volume size<br />

KSS - March 2005 - Edition 2