HHPC-2 Handheld Airborne Particle Counter Operator Manual
HHPC-2 Handheld Airborne Particle Counter Operator Manual
HHPC-2 Handheld Airborne Particle Counter Operator Manual
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
16 of 46 Setting up the <strong>HHPC</strong>-2 - <strong>HHPC</strong>-2 <strong>Handheld</strong> <strong>Particle</strong> <strong>Counter</strong><br />
<strong>Operator</strong> <strong>Manual</strong><br />
b ) Selecting Count Mode<br />
1) On the Main Screen, press the Mode key once.<br />
The bottom line displays the current mode.<br />
2) To review the modes available (listed in Table 2-1), press either the Start/Stop or the<br />
Mode key.<br />
Table 2-1 : Count Modes<br />
Fig 2-4 : Select Mode Screen<br />
Mode Description Sample Unit<br />
Concentration This mode is useful for taking a quick snapshot of<br />
airborne particulate samples, especially in areas where<br />
particulate levels are unknown and may exceed the<br />
operating levels of the counter. The <strong>HHPC</strong>-2 estimates<br />
count per cubic foot or liter depending on the volume<br />
selected. The unit updates the results by calculating<br />
completed sample counts and volume compared to the<br />
balance of volume remaining.<br />
Rate Counting is continuous. The displayed counts are a ten<br />
sample sliding average.<br />
Totalize The number of counts are displayed as they accumulate<br />
until the end of the sample. When the sample is<br />
completed, the record is stored and the value is displayed<br />
on the screen until the next sample is started.<br />
Volume size<br />
Volume unit<br />
Volume size<br />
KSS - March 2005 - Edition 2