Examples of Acceptable and Unacceptable staining with <strong>ER</strong>/<strong>PR</strong> pharmDx tM<br />
Control slides<br />
Estrogen Receptor (CAMA-1)<br />
Figure 10<br />
Unacceptable staining.<br />
Unacceptable weak staining intensity (too light);<br />
Weak staining may result in false-negative results;<br />
40x magnification<br />
Progesterone Receptor (CAMA-1)<br />
Figure 13<br />
Unacceptable staining.<br />
Unacceptable weak staining intensity (too light);<br />
Weak staining may result in false-negative results;<br />
40x magnification<br />
Figure 11<br />
Acceptable staining;<br />
40x magnification<br />
Figure 14<br />
Acceptable staining;<br />
40x magnification<br />
Figure 12<br />
Unacceptable staining.<br />
Unacceptable strong staining intensity (too dark);<br />
Excessively strong staining may result in<br />
false-positive results;<br />
40x magnification<br />
Figure 15<br />
Unacceptable staining.<br />
Unacceptable strong staining intensity (too dark);<br />
Excessively strong staining may result in<br />
false-positive results;<br />
40x magnification<br />