04.08.2013 Views

Pulsed-field gradient nuclear magnetic resonance as a tool for ...

Pulsed-field gradient nuclear magnetic resonance as a tool for ...

Pulsed-field gradient nuclear magnetic resonance as a tool for ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

214<br />

PRICE<br />

and thus, at t 2qŽ G . , with respect to<br />

the echo center, the ph<strong>as</strong>e-twisted echo will cause<br />

a coherent superposition Žwhether<br />

this is be<strong>for</strong>e<br />

or after the echo center depends on the sign of<br />

the mismatch . . Since E Ž q,. can be recovered,<br />

0<br />

it is possible to per<strong>for</strong>m signal averaging even<br />

though q and zo may fluctuate between scans.<br />

Obviously, t will vary <strong>as</strong> q fluctuates; consequently,<br />

the ph<strong>as</strong>e-twist analysis is per<strong>for</strong>med after<br />

every scan.<br />

Apart from the signal-to-noise problem, a further<br />

negative <strong>as</strong>pect of this method is that since<br />

signal acquisition occurs in the presence of a<br />

<strong>gradient</strong>, this method is not suitable <strong>for</strong> use with<br />

spectra containing more than one <strong>resonance</strong>;<br />

however, the serious <strong>gradient</strong> disturbances that<br />

warrant the use of MASSEY are normally <strong>as</strong>sociated<br />

only with me<strong>as</strong>urements of large slowly diffusing<br />

species Ž e.g., polymers . , and so spectral<br />

resolution is less likely to be an issue.<br />

SAMPLE PREPARATION AND<br />

SPECTROMETER SETUP<br />

Sample Preparation<br />

The sample and, of course, the <strong>gradient</strong> probe<br />

itself should be firmly held inside the magnet with<br />

the sample maintaining a constant position with<br />

respect to the <strong>gradient</strong> coil <strong>for</strong>mer. The sample<br />

should be wholly contained inside the linear region<br />

of the <strong>gradient</strong> coils, and thus typically the<br />

sample is contained in a volume not more than<br />

1 cm high. Such a sample, though, h<strong>as</strong> large<br />

changes in <strong>magnetic</strong> susceptibility close to the rf<br />

coils. Accordingly, it is very difficult to achieve<br />

good resolution Ž i.e., difficult to shim . . A solution<br />

is depicted in Fig. 12. This method, compared to<br />

just coaxially inserting a bulb into an NMR tube,<br />

h<strong>as</strong> an advantage in that it is e<strong>as</strong>y to clean the<br />

sample tube or work with viscous substances. It<br />

also gives a precise shape with no meniscus effect.<br />

Recently, two-piece susceptibility-matched microtubes<br />

and inserts have become commercially<br />

available. In p<strong>as</strong>sing, we note that instead of<br />

physically restricting the size of the sample, a<br />

slice-selective pulse Ž 55. could be used to restrict<br />

the sample volume. However, this is possible only<br />

on more sophisticated spectrometers, where the<br />

diffusion <strong>gradient</strong>s can be changed independently<br />

of the slice <strong>gradient</strong>. Furthermore, it should be<br />

noted that selective pulses are not of pure ph<strong>as</strong>e<br />

and do not have sharp cutoff frequencies.<br />

Figure 12 Ideally, all of the sample is contained within<br />

the constant region of the <strong>magnetic</strong> <strong>field</strong> <strong>gradient</strong> Žsee<br />

Fig. 3 . . Since the design of <strong>gradient</strong> coils are normally<br />

limited by the probe dimensions, it is generally necessary<br />

to keep the sample small so <strong>as</strong> to remain within<br />

the Ž small. volume of constant <strong>gradient</strong>. A simple solution<br />

is to place the sample in a cylindrical sample tube<br />

and then cap the sample with a vortex plug, ideally of<br />

the same <strong>magnetic</strong> susceptibility. This tube is then<br />

coaxially inserted into a tube containing either an<br />

NMR inert solvent with a similar <strong>magnetic</strong> susceptibility<br />

or the same solvent but without the solute of<br />

interest. Thus, the NMR-active part of the sample is<br />

short, where<strong>as</strong> the sample is still <strong>magnetic</strong>ally long and<br />

allows e<strong>as</strong>ier shimming. A further advantage, especially<br />

of this arrangement of the sample, is that it helps to<br />

confine the sample to the region having the most<br />

homogeneous rf.<br />

For very strong Ž in the NMR sense. samples<br />

Že.g., per<strong>for</strong>ming a diffusion me<strong>as</strong>urement on pure<br />

water . , radiation damping Ž 56. can be a problem.<br />

Since the severity of the problem is related to the<br />

magnitude of the initial magnetization, one simple<br />

solution is to use a very small sample<strong>for</strong><br />

example, by placing the sample in a small spheri-<br />

Ž .<br />

cal bulb e.g., Wilmad cat. no. 529A .<br />

B Homogeneity and Field-Frequency<br />

0<br />

Locking<br />

If the sample in question h<strong>as</strong> only one or at le<strong>as</strong>t<br />

well-separated <strong>resonance</strong>s, a high degree of resolution<br />

is of little consequence <strong>as</strong> long <strong>as</strong> there is<br />

sufficient signal-to-noise. If it is necessary to use

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!