05.08.2013 Views

Winpact Bench-Top Fermenter Instruction Manual - Wolf Laboratories

Winpact Bench-Top Fermenter Instruction Manual - Wolf Laboratories

Winpact Bench-Top Fermenter Instruction Manual - Wolf Laboratories

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Winpact</strong> <strong>Bench</strong>-<strong>Top</strong> <strong>Fermenter</strong><br />

<strong>Instruction</strong> <strong>Manual</strong><br />

FS-01-B Series(Single wall dish bottom<br />

Version: V2.4<br />

Revised on: 2012.09.13<br />

vessel)


Table of Contents<br />

www.majorsci.com<br />

service@majorsci.com<br />

Section 1 Packing list..................................................................................<br />

.. 3<br />

Section 2 Warning.................................................................................<br />

........ 5<br />

2.2Safety Information...................................................................................<br />

. 5<br />

2.3Environmental Conditions...................................................................<br />

..... 5<br />

2.4Avoiding Electrical Shock........................................................................<br />

. 6<br />

2.5Avoiding Damage to the Instrument.......................................................<br />

.. 6<br />

2.6Equipment Operation...........................................................................<br />

.... 6<br />

Section 3 Warranty Agreements and Customer Service............................<br />

8<br />

3.1Warranty agreements.............................................................................<br />

.. 8<br />

3.2Return policy.........................................................................................<br />

... 8<br />

Section 4 Introduction................................................................................<br />

. 10<br />

Section 5 Specifications...................................................................<br />

.......... 11<br />

Section 6 Delivery & Installation..........................................................<br />

...... 28<br />

2<br />

6.1Unpacking and Parts Check...................................................................<br />

28<br />

6.2Installing the fermentation system, space requirements........................<br />

28


6.3Utility Requirements.............................................................................<br />

.. 28<br />

Section 7 Component description....................................................<br />

......... 30<br />

7.1Control unit layout..............................................................................<br />

.... 30<br />

7.2Temperature control system...................................................................<br />

36<br />

7.3Culture vessel..................................................................................<br />

...... 38<br />

Section 8 Control interface......................................................................<br />

... 38<br />

1.1Monitor................................................................................................<br />

... 39<br />

8.2Pump...................................................................................................<br />

... 40<br />

8.3Control...........................................................................................<br />

........ 42<br />

8.4Setting....................................................................................................<br />

43<br />

8.5Chart.................................................................................................<br />

..... 55<br />

8.6Calibration..............................................................................................<br />

56<br />

8.7Maint. ................................................................................................ ..... 58<br />

8.8Help.....................................................................................................<br />

... 59<br />

Section 1 Operating the system..............................................................<br />

... 60<br />

1.1Preparation and connection of the control unit.......................................<br />

60<br />

9.2Prepare the Culture Vessel and Accessories.........................................<br />

61<br />

9.3Prepare the Fermentation Run...............................................................<br />

63<br />

Section 10 Remote control setting<br />

........................................................... 66<br />

3


10.1Connecting your PC to the <strong>Winpact</strong> controller.....................................<br />

66<br />

10.2Controlling the <strong>Winpact</strong> system on your PC.........................................<br />

68<br />

Section 11 Cleaning and Maintenance.......................................................<br />

72<br />

4<br />

11.1Vessel Cleaning.................................................................................<br />

... 72<br />

11.2Control Station Cleaning.............................................................<br />

......... 72<br />

11.3DO electrode information and maintenance<br />

11.4pH probe information and maintenance<br />

........................................ 72<br />

.............................................. 73


Section 1 Packing list<br />

When the unit arrives, please make sure that all items are included in the<br />

package. Use the following packing list to check the items. If an item is<br />

missing, please contact local major science representative or<br />

service@majorsci.com<br />

Item Description Qty<br />

1 Controller 1<br />

2 Culture vessel (Single wall dish bottom) 1<br />

DO/pH probe adaptor 2<br />

10mm blind Stopper (Only for 5L, 7L, 10L vessel) 3<br />

3 Brushless motor 1<br />

Accessories<br />

4 pH probe (225mm for 3L; 325mm for 5L,7L,10L) 1<br />

5 pH probe cable 1<br />

6 DO probe (220mm for 3L; 320mm for 5L,7L,10) 1<br />

7 DO probe cable 1<br />

8 Temp. probe 1<br />

9 PT-100 Temp. probe cable 1<br />

5


6<br />

10 Antifoam probe cable 1<br />

11 Power cord (EU x1, US x1) 1<br />

12 <strong>Winpact</strong> instruction manual 1<br />

Start-up kit<br />

13 250 ml glass bottle (includes 2 ports Stainless Steel<br />

Connecting set)<br />

14 500 ml glass bottle (includes 2 ports Stainless Steel<br />

Connecting set)<br />

15 45 mm 0.2mm Autoclavable disc type air filter , 10 each/pk 1<br />

16 Stainless Steel Connecting Tube, 15 ea/pk 1<br />

17 Handy Burner 1<br />

18 Silicon tubing Clamp, 12 ea/pk 1<br />

19 #16 (inner diameter 3.2mm) silicon tube 25 ft 1<br />

20 2mm Hex Wrench 1<br />

21 12mm and 14mm Double Open-End Wrench 1<br />

22 Philips (+) Screwdriver 1<br />

4<br />

1


Signed by:<br />

Date:<br />

<strong>Winpact</strong> is liable for all missing or damaged parts / accessories within 7<br />

days after customer received this instrument package. Please contact<br />

Major Science immediately regarding this issue. If no response within<br />

such time period from consignee party, that will be consignee party’s<br />

whole responsibility.<br />

7


Section 2 Warning<br />

This equipment has been tested and verified to comply with safety limits.<br />

These limits are designed to provide reasonable protection against harmful<br />

interference when the equipment is operated in a commercial environment.<br />

This equipment may generate, use, and radiate radio frequency energy, and if<br />

not installed and used in accordance with the instruction manual, may cause<br />

harmful interference with radio communications. Operation of this equipment<br />

in a residential area is likely to cause harmful interference in which case the<br />

user will be required to correct the interference at their expense. Changes or<br />

modifications not expressly approved by the party responsible for compliance<br />

could void the users authority to operate the equipment. It is strongly<br />

recommended that the user carefully read the following prints before<br />

operating any equipment.<br />

8


1.A.A.1. Read and follow the manual instructions<br />

thoroughly.<br />

1.A.A.2. Do not alter the equipment. Failure to adhere to<br />

these directions could result in personal and/or laboratory<br />

hazards, as well as invalidate equipment warranty.<br />

1.A.A.3. Use a properly grounded electrical outlet of correct<br />

voltage and current handling capacity.<br />

1.A.A.4. Disconnect from power supply before maintenance<br />

and servicing. Refer servicing to qualified personnel.<br />

1.A.A.5. In the event that solution is accidentally spilled into<br />

the instrument, disconnect grounded plug and the user<br />

must carry out appropriate decontamination. Replace<br />

damaged parts.<br />

1.A.A.6. Do not use in the presence of flammable or<br />

combustible material; fire or explosion may result. This<br />

device contains components that can ignite such<br />

materials.<br />

1.A.A.7. Refer maintenance and servicing to qualified<br />

personnel.<br />

1.A.A.8. Ensure that the system is connected to electrical<br />

service according to local and national electrical codes.<br />

Failure to properly connect may create a fire or shock<br />

hazard.<br />

1.A.A.9. Ensure the use of appropriate materials and correct<br />

operation to avoid possible hazards of explosion,<br />

implosion or release of toxic or flammable gases arising<br />

from the heated materials.<br />

9


1.A.A.10. Handle the culture vessel with cares to prevent any<br />

damages or crashes to the glass vessel, especially before<br />

and after autoclave / sterilization.<br />

1.A.A.11. Always use the supplied handle or appropriate<br />

protection to handle culture vessel in order to prevent<br />

against heat burning your hands, especially after<br />

autoclave/sterilization.<br />

ATTENTION: Hot surface!<br />

2.2 Safety Information<br />

Use high levels of precaution when using any electrical device. Before<br />

connecting the electric supply, check to see if the supply voltage is within the<br />

range stated on the rating label, and see to it that the device be seated firmly.<br />

Place the unit in a safe and dry location, it MUST NOT touch any surrounding<br />

objects. Follow the safety precautions for chemicals/dangerous materials, and<br />

hot surfaces (after autoclave/sterilization). If needed, contact a qualified<br />

service representative or service@majorsci.com.<br />

2.3 Environmental Conditions<br />

Ensure the instrument is installed and operated strictly under the following<br />

conditions:<br />

10


3.A.A.1. Indoor use only<br />

3.A.A.2. 95% RH<br />

3.A.A.3. 75 kPa 106 kPa<br />

3.A.A.4. Altitude must not exceed 2000 meters<br />

3.A.A.5. Ambient to 40°C operating temperature<br />

3.A.A.6. Pollution degree: 2<br />

3.A.A.7. Mains supply voltage fluctuations up to ±10% of the<br />

normal voltage<br />

2.4 Avoiding Electrical Shock<br />

Follow the guidelines below to ensure safe operation of the unit.<br />

The <strong>Bench</strong> <strong>Top</strong> Programmable Fermentation System has been designed for<br />

use with shielded wires thus minimizing any potential shock hazard to the<br />

user. Major Science recommends against the use of unshielded wires.<br />

11


12<br />

1.A.A.1. Dry out for a period of time and restored to<br />

NORMAL CONDITION before operation.<br />

4.A.A.1. NEVER connect or disconnect wires leading from<br />

the power jacks when the red indicator light on the<br />

Start/Stop key is on or when RUNNING is displayed on<br />

the screen.<br />

4.A.A.2. WAIT at least 5 seconds after stopping a run before<br />

handling output leads or connected apparatus.<br />

4.A.A.3. ALWAYS make sure that your hands, work area,<br />

and instruments are clean and dry before making any<br />

connections or operating the power supply.<br />

4.A.A.4. ONLY connect the power cord to a properly<br />

grounded AC outlet.


2.5 Avoiding Damage to the Instrument<br />

5.A.A.1. Do not attempt to operate the device if damaged.<br />

5.A.A.2. Protect this unit from physical damage, corrosive<br />

agents and extreme temperatures (direct sunlight etc).<br />

5.A.A.3. For proper ventilation, leave at least 10 cm of<br />

space behind the instrument, and at least 5 cm of space<br />

on each side.<br />

5.A.A.4. Do not operate the instrument in high humidity<br />

environments (> 95%), or where condensation may occur.<br />

5.A.A.5. Before using any cleaning or decontamination<br />

method except those recommended by the manufacturer,<br />

users should check with the manufacturer that the<br />

proposed method will not damage the equipment.<br />

2.6 Equipment Operation<br />

6.A.A.1. Check that the culture and all the accessories are<br />

assembled well and not damaged before starting the<br />

autoclave and fermentation process.<br />

6.A.A.2. Ensure all other associated instruments and<br />

equipment, such as the Autoclave and Cooling Water<br />

Bath are in normal condition and their specifications meet<br />

the fermentation process needs. These instruments and<br />

equipment must be operated according to their instruction<br />

manual. Furthermore, the assembly and connection<br />

between themselves and the fermentation system must<br />

be carried out correctly.<br />

6.A.A.3. NEVER access any HAZARDOUS LIVE parts that<br />

violate International Biosafety Regulations.<br />

13


6.A.A.4. When operating other fermentation systemassociated<br />

instruments, use the same level of precaution<br />

to prevent potential damage and ensure the fermentation<br />

performance.<br />

The symbols used on the <strong>Bench</strong> <strong>Top</strong> Programmable Fermentation<br />

System are explained below.<br />

14<br />

Indicates an area where a potential shock hazard may<br />

exist.<br />

Indicates a warning. Consult the manual to avoid<br />

possible personal injury or instrument damage.<br />

ATTENTION: Hot surface!


Section 3 Warranty Agreements and<br />

Customer Service<br />

3.1 Warranty agreements<br />

Major Science warrants its products against defects in materials and<br />

workmanship, under normal service, for a period of 12 months from the<br />

shipping date to purchaser, unless other terms are agreed upon in writing.<br />

The date of reference is the date of the delivery, which must be evidenced by<br />

presenting the corresponding delivery confirmation.<br />

The guarantee is given against defects in manufacture and against<br />

malfunctioning, but not for parts subject to wear and tear (O-ring, seals,<br />

membrane filter), defects or damages caused by improper handling, or<br />

defects and malfunctions caused by corrosion (due to lack of resistance<br />

against solvents, etc., which are used for fermentation).<br />

The warranty does not cover parts subject to wear and tear or damages<br />

caused by improper handling. In addition, it does not cover the following:<br />

15


1.A.A.1. Corrosion/damage to the tubing clamps on the<br />

peristaltic pump, which is caused by contact with strong<br />

acid/base or chemicals<br />

1.A.A.2. Damage to the fermenter controller, vessel and<br />

cables caused by negligence during installation or<br />

adjustment<br />

1.A.A.3. All probes (pH, DO; excluding DO probe<br />

membrane) are covered by the warranty for six months<br />

starting from date of completion of installation<br />

1.A.A.4. O-rings, mechanical seals and start-up kits<br />

components (silicone tubing, clamp, feeding bottles, etc.)<br />

1.A.A.5. All glass parts are excluded from the warranty<br />

Note: Before first use or use under specific environmental<br />

conditions in the laboratory and/or together with special<br />

nutrients or additional solutions, you must test the resistance of<br />

all parts.<br />

Major Sciences obligation under this warranty is limited to repairing parts or<br />

providing replacement parts at no charge, which prove to be defective during<br />

the warranty period. A part shall be considered defective after inspection of<br />

Major Sciences technical staff. At Major Sciences option, we will repair or<br />

replace any defective part, which is returned to our facilities. The cost of<br />

shipping the repaired or replacement part will be borne by Major Science. Any<br />

unit where repair or modification has been performed by anyone other than<br />

Major Science or an appointed distributor/representative is no longer under<br />

warranty from the time the unit was modified.<br />

3.2 Return policy<br />

A local Major Science representative can repair defects and damages.<br />

Defective appliances can also be returned to a local Major Science<br />

16


epresentative. It will be the customers liability to pay all carriage charges.<br />

The repairs will be carried out and charged in accordance with our terms of<br />

maintenance.<br />

Note: Upon return, the equipment must be clean, in good<br />

hygienic condition and carefully packed. Contagious parts must<br />

be disinfected or sterilized, according to chemical, biological,<br />

biotechnological or genetic safety regulations. The sender has to<br />

prove compliance with corresponding safety regulations. The<br />

sender will be charged for repair or damages due to transport<br />

and for necessary cleaning and disinfection of any parts.<br />

For returning any equipment, please notify your local major science<br />

representative or contact Major Science directly:<br />

No. 37, Wuquan 5th Rd., Wugu Dist., New Taipei City 24888 Taiwan, R.O.C.<br />

TEL: 886-2-2298-1055<br />

FAX: 886-2-2299-7871<br />

19959 Sea Gull Way, Saratoga, CA 95070, U.S.A.<br />

TEL:1-408-366-9866<br />

FAX:1-408-446-1107<br />

E-mail: service@majorsci.com<br />

17


Section 4 Introduction<br />

<strong>Winpact</strong> Scientific is a product brand under Major Science, an innovative,<br />

R&D-based<br />

manufacturer supplying a broad product portfolio to the life science market.<br />

<strong>Winpact</strong> provides a comprehensive and innovative line of cultivation products<br />

designed for different cell culture experiments and applications. It comes in a<br />

benchtop scale and has a large, color touch-screen panel with a user-friendly<br />

interface. Its distinctive functions include various programming operations to<br />

control the pump speed, pH levels, temperature, and more. <strong>Winpact</strong><br />

Fermentation System is equipped with a full connection device to connect to<br />

any PC for real-time recording and control within the vessel<br />

Features<br />

18<br />

-104 color touch screen & graphical user interface<br />

-Ethernet remote control and data logging function<br />

-Flexible optional device selections<br />

-Storage program: Up to 59,994 programs for different kind<br />

of conditions<br />

-Data storage: Up to 100 data files<br />

-Data export interface: USB port<br />

-Modular system for easy upgrade & maintenance<br />

-Full accessory as a package


Section 5 Specifications<br />

Vessel Max.<br />

Working<br />

volume<br />

Total<br />

volume(Liter<br />

)<br />

Inner<br />

diameter(m<br />

m)<br />

Inner<br />

height(mm)<br />

Overall<br />

diameter(m<br />

m)<br />

Height without<br />

condenser(mm)<br />

Height with<br />

condenser(mm)<br />

Single wall (Dish bottom)<br />

3L 5L 7L 10L<br />

3.8 6.8 8.9 12.5<br />

130 160 180 190<br />

310 365 370 430<br />

200 230 250 270<br />

520 570 580 630<br />

640 640 650 700<br />

Material Borosilicate glass / 316L stainless steel for<br />

headplate and all fittings<br />

19


Control unit Control panel 10.4 Color Touch-screen Interface<br />

20<br />

Aeration Inlet Gas<br />

Temperatur<br />

e<br />

(Resolution: 800 x 600 pixels)<br />

Communication port Remote control through Ethernet<br />

Storage program Up to 59,994 programs for different kinds<br />

of condition<br />

Data storage Up to 100 data files<br />

Data storage<br />

interface<br />

USB port<br />

Cabinet material ABS front panel and painted iron housing<br />

Dimension Footprint: 400 x 500 mm (W x D) ; Height:<br />

740 mm<br />

Rated voltage 110V or 220V ; 50/60 Hz<br />

Flow-meter<br />

0, 1 10<br />

LPM<br />

Sparger Orifice ring<br />

0, 2 20 LPM<br />

Baffle Removable 316L stainless steel baffles<br />

Heating Built-in heat exchanger<br />

Cooling Cooling coil valve<br />

Range 5 ℃ above coolant up to 60℃<br />

Probe Platinum RTD probe (Pt-100)<br />

Control mode Programmable 15 steps PID controller


Agitation Drive Removable <strong>Top</strong> brushless motor<br />

Speed Range 30 1200 rpm (3L~7L) ; 30-1000rpm<br />

(10L)<br />

Impeller Three adjustable Rushton-type impellers<br />

(standard);<br />

Two Pitched-blade impellers (optional)<br />

Control mode Programmable 15 steps PID controller<br />

pH Range 2 -14 pH<br />

Probe Gel-filled electrode; Autoclavable<br />

Control mode PID, one point control with adjustable<br />

deadband<br />

DO Range 0 200%<br />

ORP(Option<br />

al)<br />

Probe Polarographic DO sensor; Autoclavable<br />

Control mode PID; Cascade function to response to<br />

Transmitter -/+ 2000 mV<br />

Resolution 1 mV<br />

a. Increase or decrease agitation speed<br />

b. Oxygen Enrichment module (Optional)<br />

c. Gas Mixing Station module (Optional)<br />

Start substrate feeding program<br />

Probe Gel-filled electrode; Autoclavable<br />

Foam Probe 316L stainless steel detector with<br />

insulated teflon tube; On/Off control<br />

21


Peristaltic<br />

pump<br />

Pump number Four built-in pumps; One external pump<br />

(Optional)<br />

Motor type Precise stepping motor; minimum speed<br />

is 1 rpm<br />

Speed range 0 65 rpm<br />

Control mode Programmable 15 steps feeding control;<br />

Pump can be assigned for Acid, Base,<br />

Antifoam and Substrate<br />

Exhaust Device type 316L Stainless steel condenser<br />

Utility<br />

requirement<br />

22<br />

Power supply 100-120V 50/60 Hz or 210-230V 50/60 Hz<br />

with electrical safety cutoff switch<br />

Water 2 Bar maximum (29 psi)<br />

Air 1 Bar maximum, must be dry, oil-free and<br />

filtered<br />

Autoclave For sterilization


Section 6 Delivery & Installation<br />

6.1 Unpacking and Parts Check<br />

Please pay attention to the following advice:<br />

1.A.A.1. Carefully check the outer packaging, the packaging<br />

of the single components, as well as the parts themselves<br />

for damage.<br />

Warning: Do not use culture vessel when damage is suspected.<br />

In such a case, the glass vessel might burst during sterilization<br />

in the autoclave.<br />

1.A.A.2. Check if the delivery is complete. To prevent<br />

scratching, please handle with care when using sharp<br />

objects to unwrap the packaging.<br />

1.A.A.3. Please use caution when loading or unloading to<br />

prevent damage due to collision.<br />

1.A.A.4. When moving or positioning the controller, take<br />

care not to lift it by holding the peristaltic pump heads.<br />

They are not designed to be used as handles, and can be<br />

damaged by lifting a heavy load.<br />

1.A.A.5. If parts are missing or damage occurs in transit,<br />

please inform your local Major Science representative.<br />

23


24<br />

6.2 Installing the fermentation system, space requirements<br />

2.A.A.1. The system will require an area of approximately W<br />

x D = 1,500 x 600 mm (including the space for the culture<br />

vessel and the cooling circulating water bath)<br />

2.A.A.2. You may vary the arrangement of the components<br />

according to the lengths of the connecting lines. It is<br />

recommended to consider some additional space for<br />

convenient handling of the equipment.<br />

2.A.A.3. Use the (+) Screwdriver to open the back panel for<br />

gas pressure adjustment before first use (Screwdriver is<br />

provided in the standard package). Detailed instructions<br />

are given below.<br />

6.3 Utility Requirements<br />

3.A.A.1. Gas Source: the incoming gas must be reduced to<br />

at least 2 bar (29psi) before entering the air inlet (Fig. 1).<br />

If you are using an air compressor or a cylinder, please be<br />

sure to use an external regulator to reduce the gas<br />

pressure. The inlet gas must be further adjusted to 0.5 bar<br />

(approx. 7psi) using the built-in pressure gauge (Fig. 2),<br />

the back panel must be opened for access). The built-in<br />

pressure gauge will be fully shut off when the<br />

fermentation system first arrives, which means no air flow<br />

will be observed in the rotameter. Open the back panel<br />

and lift up the pressure gauge knob to adjust. Turn<br />

counterclockwise to open the pressure valve and turn<br />

clockwise to close the pressure valve. After adjusting to<br />

0.5 bar, push the knob down to lock and close the back<br />

panel.


Fig. 1 Fig. 2<br />

3.A.A.2. Coolant Source: Please make sure your water<br />

quality is clean (50 micron filter). The hardness of water<br />

should be below 12 German Degree of Hardness in order<br />

to prevent calcareous deposits. The water pressure must<br />

be reduced to 1 bar (14.5psi) to prevent damage to the<br />

valve. To achieve a good cooling efficiency, we<br />

recommend setting the coolant temp. at least 15 20 º C<br />

below the desired culture temp. Using tap water is not<br />

recommended but workable. Please tighten the main<br />

water inlet connection to prevent leaking or bursting.<br />

Damage caused by impure water quality will not be<br />

covered by the warranty.<br />

Suggested tubing size for coolant & air connection (Fittings might be needed<br />

for connecting to your air & coolant source):<br />

For chiller (Main water inlet & Main water outlet port): 3/8 (9.6mm)<br />

For Air (Gas outlet, Air inlet & O2 inlet): 3/16 (4mm)<br />

25


Section 7 Component description<br />

The system is comprised of three main functional units: The Control unit, the<br />

Culture vessel, and the Heating blanket / Heating base unit. This chapter<br />

introduces the functions of each unit and gives general guidance for the<br />

machine operation.<br />

7.1 Control unit layout<br />

No. Description Function<br />

26<br />

1 10.4 color touch screen Graphical interface (Resolution: 800x600<br />

pixels)


2 Power switch ON/OFF Switch for the controller<br />

3 Air Rotameter Regulate the air flow<br />

4 Peristaltic Pump Control pH, foam level and feed nutrients<br />

5 pH probe Connect to pH probe<br />

6 ORP probe Connect to ORP sensor<br />

7 Antifoam probe Connect to antifoam sensor<br />

8 Temp. probe 2 Connect with blanket temp. probe / Measure<br />

the blanket temp.<br />

9 Vessel Temp. sensor Connect to the vessel temp. sensor (PT-100)<br />

10 GAS mixing Connect to the gas mixing station (Optional)<br />

11 Motor Connect to the brushless motor<br />

12 Base unit/Blanket Perform heating function<br />

13 DO probe Connect to the DO probe<br />

14 Condenser in Connect to the condenser inlet<br />

15 Condenser regulator Control the water flow of the condenser<br />

16 Main water in Coolant inlet<br />

17 Main water out (drain) Coolant return<br />

18 Gas out Connect to the air sparger<br />

19 Condenser out Connect to the condenser outlet<br />

27


20 Water out Connect to cooling coil (Jacket) outlet<br />

28<br />

21 Water in Cooling to cooling coil (Jacket) inlet<br />

22 O2 in Connect to the oxygen source (Optional)<br />

23 Air in Connect to the air source (External air pump or<br />

house air)<br />

24 USB port For data export<br />

25 Power socket Power supply<br />

26 Ethernet port Connect to the Ethernet and perform remote<br />

control function<br />

27 RS-485 For system expansion<br />

28 Pump 5 For external 5th pump<br />

29 AUX 1 For system expansion<br />

30 AUX 2<br />

31 O2 flowmeter (Optional) Control the oxygen flow (Only available with O2<br />

enrichment purchase)


7.2 Temperature control system<br />

2.A. Cooling system<br />

Please go to Help page to check the detailed connection instructions.<br />

No. Description Function<br />

1 Condenser in Connect to the vessel condenser inlet<br />

2 Condenser out Connect to the vessel condenser outlet<br />

3 Condenser regulator Control the water flow of the condenser<br />

4 Main water In Connect to the coolant inlet<br />

5 Main water out<br />

(Drain)<br />

Connect to the coolant return<br />

6 Water out Connect to cooling coil (Jacket) outlet<br />

7 Water in Cooling to cooling coil (Jacket) inlet<br />

29


8 O2 in Connect to the oxygen source<br />

9 Air in Connect to the air source<br />

10 Gas out Connect to air sparger port on the headplate<br />

Note:<br />

The condenser water flow is controlled by a condenser water flow regulator<br />

(See figure below). The cooling coil water flow is controlled by a solenoid<br />

valve. The solenoid valve opens only when the vessel requires cooling<br />

function. When performing temp. control, the valve will open to let water flow<br />

into the cooling coil. At the same time, you will also hear the sound of the<br />

solenoid valve, which means the valve is functioning correctly.<br />

In order to provide constant water flow through the condenser, we must turn<br />

the regulator knob to open the regulator valve halfway. The purpose is to<br />

prevent media evaporation. The condenser helps evaporated media re-<br />

condense back to liquid.<br />

30


2.B. Heating systems<br />

No. Name Description<br />

1 Coolant valve Control the coolant inlet flow<br />

2 Heat exchanger Built-in heater to provide heating function<br />

3 Circulation pump Help water circulate in the vessel<br />

31


32<br />

2.B.A.1. The minimum operating temperature in the culture<br />

vessel will be 5 ℃ above the coolant. To provide an<br />

efficient cooling control, the chiller temp. is normally set at<br />

least 10 ℃ below the culturing temperature.<br />

2.B.A.2. The water must be clean and free of particles.<br />

2.B.A.3. The hardness of water should be below 12 German<br />

Degree of Hardness in order to prevent calcareous<br />

deposits.<br />

Defects and damage caused by dirty water or calcareous<br />

deposits will not be covered by our warranty!


7.3 Culture vessel<br />

No<br />

.<br />

1<br />

2<br />

3<br />

4<br />

5<br />

3.A. Headplate Arrangement<br />

Feeding port<br />

Inoculation port<br />

To air sparger<br />

Septum port<br />

AntiFoam probe<br />

Name Function<br />

For Acid/Base/Antifoam/Substrate addition<br />

To introduce the inoculum<br />

Introduce air into the culture media<br />

Small volume sample injection<br />

Detect the foam level<br />

33


6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

13<br />

34<br />

Spare port for DO/pH probe<br />

Thermowell<br />

Cooling coil in<br />

Cooling coil out<br />

Condenser<br />

Sampling port<br />

T-hand lifter<br />

Screw nut<br />

3.B. Vessel body<br />

For installing the DO/pH probe<br />

For Pt-100 temp. probe insertion<br />

Connect to Water in port on the controller<br />

Connect to the Water out port on the<br />

controller<br />

Prevent culture media evaporation<br />

Sampling/harvesting the culture media<br />

For vessel transfer<br />

For headplate fixation


No<br />

.<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

Glass body<br />

Headplate<br />

Supporting rods<br />

Name Function<br />

Impeller(Pitched-blade type)<br />

Sparger<br />

Baffle<br />

Condenser in port<br />

Condenser out port<br />

Made from borosilicate glass<br />

Arrange each port on the top plate<br />

Support/Protection for the vessel body<br />

Perform mixing and agitation<br />

Sparge gas into the culture vessel<br />

Disturb vortex and increase aeration<br />

efficiency<br />

Condenser coolant inlet<br />

Condenser coolant outlet<br />

Section 8 Control interface<br />

The control interface can be structured into 8 display screens depending on<br />

the functions:<br />

35


1.A.A.1. Monitor: Shows the status and readings from the<br />

sensor<br />

1.A.A.2. Pump: Controls the peristaltic pump speed and<br />

movement<br />

1.A.A.3. Control: Performs manual or automatic control of<br />

each parameter<br />

1.A.A.4. Setting: Sets up a schedule for the automatic<br />

process<br />

1.A.A.5. Chart: Plots and records the trend of a<br />

fermentation process<br />

1.A.A.6. Calibration: Performs sensor calibration<br />

1.A.A.7. Maint.: For service and maintenance only<br />

1.A.A.8. Help: <strong>Instruction</strong> for cooling system and tubing<br />

connections<br />

1.1 Monitor<br />

Once you start the system, monitor page will be shown as the first screen.<br />

Detailed descriptions are classified into nine areas (Area A - I).<br />

36


Area A: Indicates the system condition. If this area shows a green light and<br />

ON-LINE, the system is working under normal condition. If it shows a red<br />

light and OFF-LINE, there may be a problem, and you should consult with a<br />

<strong>Winpact</strong> service operator<br />

Area B: This area allows you to switch between each page<br />

Area C: Displays the readings from sensors such as DO (Dissolved oxygen),<br />

pH, Temp. and optional devices such as ORP or CO2/O2 Analyzer. Also<br />

shows the set value for your control process (SV: set value; PV: present<br />

value). If the optional device is selected, the readings will also be displayed<br />

(Ex. CO2/O2, O2 Flow)<br />

Area D: System login button. Provides access for different users<br />

Area E: Displays the condition of each device. If the device is currently<br />

operating, the color will change to flashing red<br />

Circulation pump Solenoid valve Heat exchanger<br />

37


Area F: When the system is logged in via remote control, this area will display<br />

the remote IP address<br />

Area G: Timer for process time calculation<br />

Area H: Displays the system operation mode. In this case, the screen display<br />

is MANUAL to indicate the system is operating under manual mode.<br />

Area I: This area shows the status of the fermentation process. When<br />

entering the automatic mode from the manual mode, the message will switch<br />

from the ellipsis to the running table message to indicate the automatic<br />

process.<br />

Note 1: If your system is equipped with the O2 enrichment device (optional),<br />

your monitor screen will show an additional O2 solenoid valve icon (Boxed<br />

area). You can use this to manually control oxygen flow.<br />

Note 2: Your vessel display will change according to your System Setup<br />

selection on Maintenance page (please see section 8.7)<br />

38


1.B. How to control the oxygen gas flow(Optional device)<br />

1.B.A.1. Connect your oxygen source to the O2 in port on<br />

the controller. Please go to Section 6.3 for more details.<br />

1.B.A.2. Open the controller back panel and you will find<br />

two pressure gauges (see figure below). The upper one is<br />

for oxygen, lower one is for air. Open the oxygen source<br />

and manually adjust the pressure gauge to the desired<br />

value. The pressure limit for the culture vessel 1 bar (14.5<br />

psi). If you are using a cylinder as the O2 source, please<br />

be sure to use an external regulator to reduce your inlet<br />

gas pressure (see Section 6.3).<br />

1.B.A.3. On the front panel, press the button to<br />

manually open the oxygen valve. Adjust the flow rate<br />

using the oxygen rotameter.<br />

39


40<br />

1.C. How to login to the system<br />

1.C.A.1. Please login before you start system operation<br />

1.C.A.2. Press and the login dialog will pop out<br />

1.C.A.3. Select Supervisor and key in 1234 (Default<br />

password) as the password<br />

1.C.A.4. Press OK then system will enter the monitor page<br />

1.C.A.5. To log out of the system, press<br />

Note: Users can access the system using different user names<br />

and user levels. Factory mode is for manufacturer and system<br />

maintenance purposes. The user can only choose “ Supervisor”<br />

and “ User” as the user ID. The difference between these two user<br />

ID’s will be explained in detail in Section 8.4. To change the<br />

password, also see Section 8.4.


8.2 Pump<br />

2.A.A.1. How to change the pump speed:<br />

Assign pump speed from the rpm column<br />

Peristaltic pump speed range: 1 65 rpm. Increment: 1 rpm<br />

2.A.A.2. How to change the pump direction<br />

Use and to switch the pump direction forward or backward.<br />

41


2.A.A.3. Switch between manual/automatic mode<br />

Use to switch between <strong>Manual</strong> and Automatic control.<br />

2.A.A.4. Pump 5 is for the optional external pump<br />

Four silicon tubing sizes are available. Please check the table<br />

below for the nominal flow rate.<br />

Nominal flow rate at 1 rpm (ml/min)<br />

Tube inner<br />

diameter<br />

0.8mm 1.6mm 3.2mm 4.8mm<br />

Flow rate at 1 rpm 0.06 0.21 0.8 2.8mm<br />

2.A.A.5. How to turn off the peristaltic pump<br />

In automatic mode:<br />

Go to setting page, press and enter the pump application page, choose<br />

which pump you want to switch off and switch the pump application to .<br />

This assigns the specific pump to no application.<br />

42


8.3 Control<br />

You can either manually control each parameter or run automatic control of<br />

your fermentation process<br />

Area A: Automatic mode<br />

3.A.A.1. Start an automatic fermentation process<br />

Press the button. The system will start running the process<br />

according to the Set Temperature / Stirring table and pump-feeding schedule.<br />

To set up the table and schedule, go to the Section 8.4 setting page.<br />

3.A.A.2. ACID Start function<br />

If you use the function, the automatic mode will only be<br />

43


triggered when the pH reaches the set value. To set the pH start point, go to<br />

the Section 8.4 setting page for Acid-Start setup.<br />

Note: This function is specifically designed for lactobacillus<br />

culture. The cells produce acidic metabolite as they grow and the<br />

pH drops continuously. When the pH drops to the set points, the<br />

automatic mode will activate and the whole operation turns to<br />

automatic mode by running the rpm table, Temp. table, Feeding<br />

table and controls the pH within the set range (pH Auto-setup).<br />

3.A.A.3. Stop an automatic fermentation process<br />

Press to terminate an automatic fermentation process.<br />

Area B: <strong>Manual</strong> control mode<br />

44<br />

1.A.A.1. Agitation control<br />

-Change the agitation speed from the RPM column.<br />

-Press to start manual agitation.<br />

-To stop agitation, press again.


3.A.A.4. Temperature control<br />

-Change the temp. from the temp. column<br />

.<br />

-Press to perform temp. control<br />

-Press to disable temp. control<br />

3.A.A.5. pH control parameter Setting<br />

-Change the pH value from the pH column<br />

-Press to start pH control<br />

-Press to stop pH control<br />

Note: Before using the pH control function, please assign pump<br />

applications for pumping Acid or Base. You can also define the<br />

dead-band (pH control range) and pulse interval for pumping the<br />

reagents. The detailed settings will be mentioned in Section 8.7<br />

maintenance page.<br />

45


3.A.A.6. DO Cascade<br />

-Press to start DO cascade control.<br />

-Before using the DO cascade function, go to the setting<br />

page to change the DO cascade parameters. The detailed<br />

setting will be mentioned in the next section.<br />

8.4 Setting<br />

In this page, you can program the automatic fermentation process and set up<br />

the DO cascade parameters<br />

46


4.A. Automatic Agitation Program<br />

Press to program the automatic agitation schedule.<br />

Fill in the table according to the desired settings.<br />

Note: You can set up 15 steps. In this case, the first step will run<br />

at 25 rpm for 60min., then 200rpm for next 60min, and so on.<br />

Note: To fully stop agitation under programmable mode, you will<br />

need to set two step settings. For the first step, set RPM to 0, and<br />

MIN. to 1, then on the second step, set RPM to 0 and MIN. to 0<br />

(The program stops at the previous step indefinitely, which in<br />

this case would be 0 RPM).<br />

47


4.B. Automatic Temperature Program<br />

Press to program temp. setting. The process of the temperature<br />

program is basically the same as agitation program.<br />

Note: There are 15 steps available. In this case, the temp. is kept at 50 ℃ for<br />

the first 60 mins, 60 ℃ for the next 60 mins, and so on.<br />

Note: You cannot stop temperature control fully under programmable mode.<br />

48


4.C. Peristaltic Pump setting page<br />

4.C.A.1. Press and enter the pump application page.<br />

Assign an application for each peristaltic pump. You can<br />

assign the peristaltic pump to pump Acid, Base, Antifoam<br />

or Substrate. You can also assign one to four pumps for<br />

the substrate feeding:<br />

Acid:<br />

Base:<br />

Antifoam:<br />

Substrate feeding:<br />

Note: When using CO2 gas as acid control, acid pump will be disabled.<br />

49


Example:<br />

4.C.A.2. To set up a feeding schedule, use the table above<br />

to run the specific schedule you need. You can set up the<br />

pulse interval using the table.<br />

If you need to program Feeding for continuously feeding for 2 hours, you<br />

could set the feeding pump ON for 60 seconds, OFF for 0 seconds and 120<br />

cycles in step 1; or set ON for 60 minutes, OFF for 0 minutes and 2 cycles in<br />

step 1.<br />

In this page, pump 1 is assigned for pumping Feed 1, pump 2 for Antifoam,<br />

pump 3 for Base and pump 4 for Acid. The feeding pump is set at ON for 5<br />

seconds, OFF for 2 seconds and 10 cycles in step 1. When finishing step 1,<br />

the feeding pump will start entering step 2. Up to 15 steps are available in this<br />

table.<br />

You can set up the feeding schedule based on your tubing flow rate. With this<br />

table, you can calculate how much volume can be pumped into the vessels.<br />

You can change the time unit by switching the Second/Minute button<br />

50<br />

4.C.A.3. To cancel pump applications, switch the pump<br />

application to<br />

4.D. DO cascade Setting<br />

4.D.A.1. DO (Dissolved Oxygen) Cascade<br />

Our DO cascade program is specifically designed to meet most of your<br />

experimental needs when running a fermentation/bio-reaction experiment.<br />

DO control usually does not come online until bacteria/yeast starts to grow<br />

exponentially, which consumes tremendous oxygen in a short period of<br />

time.<br />

So the purpose of the DO Cascade program is to maintain the DO level<br />

within the system once the DO set point value has dropped to a significant


level which may affect the result of the experiment. If DO level never<br />

reaches the set value, the DO cascade program will not initiate hence it<br />

will not be activated.<br />

For example, if your DO set point (SV) is set at 40% and the present value<br />

(PV) is 65%, DO cascade will not activate since PV has not dropped below<br />

40%. No action will be taken by the DO Cascade program because it has<br />

not been triggered or activated.<br />

4.D.A.2. DO Set point<br />

DO set point acts as an activator for the DO cascade program (as<br />

explained in the previous section) and also acts as a control point for your<br />

DO level.<br />

This is the set value you require for the DO level during your experiment.<br />

For example, DO level is set at 40.0%, which means when DO Cascade is<br />

triggered, it will maintain DO level at 40.0%.<br />

This setting here allows you to add substrates through your pump feeding<br />

when DO level drops below the set value (e.g. 40.0%). You could select<br />

the feeding pump you assigned on (using the pump setting). Once the<br />

feeding program is triggered, it will run to completion.<br />

4.D.A.3. Setting Menu<br />

51


The DO Cascade setting menu is located under the Setting page. Its<br />

feature can be divided into two sections: Control Strategy and Feeding<br />

Strategy.<br />

Control Strategy (1):<br />

Settings in the black box, this strategy accommodates majority of the user<br />

needs. These settings allow you to adjust DO level using Agitation, O2<br />

enrichment module, O2 with mass flow controller enrichment or Gas<br />

Mixing Station.<br />

Feeding Strategy (2):<br />

52<br />

Settings in the red box, this is commonly used for any user who wants to<br />

harvest secondary metabolites during the experiment or needs to activate<br />

a feeding program at a specific DO condition. Hence with the First Trigger<br />

and Second Trigger options, whereas Second Trigger will not activate


unless First Trigger condition has been met and is true. Details of this<br />

strategy are explained on page 7.<br />

4.D.A.4. Control Strategy<br />

After setting your DO value, lets continue to the right side of the screen,<br />

you should see there are 2 stages available for your DO control. When DO<br />

Cascade is in motion, the program will run stage 1 first, if stage 1 method<br />

fails to maintain the DO level, the system will proceed to stage 2.<br />

Selecting the right method to use:<br />

By pressing the rectangular button, you could select which specific method<br />

you prefer to run for each stage.<br />

Different Methods and related input values:<br />

b.1 Stirr<br />

Select this to use agitation to increase or decrease your DO<br />

level<br />

53


54<br />

Time (sec): duration of each cycle<br />

Step (rpm): this is the rpm value for each increment or<br />

decrement during each cycle<br />

Max (rpm): this is the maximum rpm value for this stirring<br />

method<br />

Min (rpm): this is the minimum rpm value for this stirring<br />

method<br />

b.2 Built-in O2<br />

b.3 Built-in O2<br />

b.4 O2 (GM)<br />

Select this when connecting with MS O2 enrichment module<br />

(FS-O-OE). The cycle duration will follow your O2 setting from<br />

the Maintenance page (please see page 9 for the O2 setting).<br />

All values here are irrelevant.<br />

Select this when connecting with MS Oxygen enrichment with<br />

mass flow controller (FS-O-MF)<br />

Time (sec): the calculation cycle of each time interval for the<br />

increment/decrement, sampling time<br />

Step (%): the % of the increment/decrement of the total time<br />

interval, T1 (which could be adjusted in the<br />

maintenance page)<br />

Max (%): the maximum amount of time for the solenoid<br />

valve to be opened<br />

Min (%): the minimum amount of time for the solenoid valve<br />

to be opened<br />

Select this when connecting with MS Gas Mixing Station (FS-<br />

O-GM) and would like to adjust DO level using O2 only.<br />

Time (sec): the calculation cycle of each time interval for the<br />

increment/decrement<br />

Step (%): the % of the increment/decrement of the total time<br />

interval, T1 (which could be adjusted in the


maintenance page)<br />

Max (%): the maximum amount of time for the solenoid<br />

valve to be opened<br />

Min (%): the minimum amount of time for the solenoid valve<br />

to be opened<br />

b.5 O2/N2 (GM): (FS-O-GM)<br />

Select this when connecting with MS Gas Mixing Station (FS-<br />

O-GM) and would like to adjust DO level using O2 for<br />

increasing DO and N2 for decreasing DO.<br />

Time (sec): the calculation cycle of each time interval for the<br />

increment/decrement<br />

Step (%): the % of the increment/decrement of the total time<br />

interval, T1 (which could be adjusted in the<br />

maintenance page)<br />

Max (%): the maximum amount of time for the solenoid<br />

valve to be opened<br />

Min (%): the minimum amount of time for the solenoid valve<br />

to be opened<br />

Operating procedure using Control Strategy<br />

The control strategy allows you to maintain your DO using the two<br />

methods of your choice.<br />

For the setting given above, stage 1 is using the Stirr method and stage<br />

2 is using the Built-in O2 method. Once the DO level drops below 40%,<br />

55


56<br />

DO Cascade will activate the Stirr method by increasing the agitation<br />

speed by 20 rpm per cycle. If the current setting is 200, for every 5<br />

seconds, it will increase 20 rpm to 220 rpm, and check the DO level<br />

again, if another increment is required, it will increase another 20 rpm<br />

to 240 by the next 5 seconds, and if the system reaches your set max<br />

RPM value (in this case, 500 rpm) and still unable to adjust the DO<br />

level to 40%, it will proceed to Stage 2.<br />

For Stage 2, as the example given below, this setting is using the Built-<br />

in O2 enrichment module, as previous stated, if you have purchased<br />

Gas Mixing station, Oxygen enrichment with mass flow controller, or<br />

simply the oxygen enrichment module, you need to select it<br />

appropriately in order for the setting to work properly.<br />

c.1 Flow chart of increasing DO


c.2 Flow chart of decreasing DO<br />

4.D.A.5. Feeding Strategy<br />

Once your DO level falls/reaches your set value (First Trigger value, in %),<br />

it will activate your second Trigger Value, once the DO raises up to the set<br />

value, the feeding schedule (by selecting the Feed you want) will initiate.<br />

The feeding strategy is designed for experienced user who understands<br />

his experience thoroughly and would like to schedule a feeding schedule<br />

based on the metabolic process of his bacteria such as lactobacillus. This<br />

feature allows you to add substrates feeding based on the changes of the<br />

DO level.<br />

Operating procedure using Feeding Strategy (red box)<br />

57


58<br />

As the setting given above, when the DO level drops from 40% to 20%,<br />

it will active the first trigger, which is still in idle mode but waiting for the<br />

second trigger value to be active. Once the DO level rises from 20% to<br />

50%, if a feeding schedule is selected, it will be activate.<br />

Note: if DO level rise directly from 40% to 50%, it will not activate the<br />

feeding strategy because the first trigger value (20%) was not reached.<br />

4.D.A.6. Finding your T1 (total time interval)<br />

Go to Maintenance page, on this page you should be able to locate the<br />

total time interval for the oxygen valve. (for FS-O-OE and FS-O-MF)


T1 = the cycle duration in time for the oxygen valve<br />

T2 = the duration in time the valve will open during each cycle.<br />

Note: when selecting your O2 source, these values (values in blue box<br />

from the picture below) will not apply to O2 enrichment module. Instead,<br />

Cascade will only use the pulse interval of the oxygen cycle (T1 and T2 in<br />

red box from the picture above) to adjust your DO level. All values are<br />

irrelevant.<br />

59


60<br />

Example: In the settings given below:<br />

Once the first stage of the DO cascade has failed to maintain the dissolved<br />

oxygen level, the system will automatically enters the second stage and<br />

introduce pure oxygen to raise the DO value. Make sure to have the<br />

Oxygen selection set to ON in order to have this feature to work properly.<br />

In this case, the Oxygen valve will open every 10 seconds, with increment<br />

of 1 second (10% of 10 seconds [T1] = 1 second) with maximum opening<br />

valve time as 10 seconds (100% of 10 seconds [T1] = 10 seconds) and<br />

minimum opening valve time as 1 second (10% of 10 seconds [T1] = 1<br />

second).<br />

So the system will start with opening valve for 1 second, and close for 9<br />

seconds, if it failed to maintain the DO level, it will proceed with opening<br />

valve for 2 second (1 second initial with increment of 1 second), and close<br />

for 8 seconds, and so on.<br />

For every 10 seconds (Time), the system will monitor the DO level to<br />

calculate the increment or decrement of the solenoid valve.


4.D.A.7. The Restore Delay<br />

The minimum duration in time (only in digits, not decimal points) for which<br />

the system will evaluate before it de-activate the program. (This is to make<br />

sure DO level reaches the set point and is stabilized)<br />

4.D.A.8. The Response dead-band value<br />

As our systems standard value, the response dead-band is 5 %. This<br />

value is the neutral buffer area where no action occurs. The purpose of the<br />

response dead-band value is to prevent oscillation or repeated activation-<br />

deactivation cycles of DO cascade, which may cause great fluctuation of<br />

the DO level and shorten the life of the solenoid valve. This means that if<br />

your DO set point is at 40%, the DO cascade program will not initiate and<br />

until the DO level reaches 35% (40 % - 5% [dead-band value]); vice-versa<br />

the program will sustain in the idle when the DO level reaches 45% (40% +<br />

5% [dead-band value]).<br />

Note: Response dead-band could be adjusted according to your need,<br />

61


minimum setting is 1.5 %.<br />

4.E. File Management<br />

We use different parameters for different growth conditions. To save and<br />

manage these files, you can use the New/Open/Copy/Rename/Delete<br />

function.<br />

62<br />

4.E.A.1. Open a new file: Press the New button then the<br />

following dialog will show up. Choose a file name title<br />

(LAB1 ~ LAB5) and key in a number (0001 ~ 9999). Then<br />

press OK to establish a new file. You can save up to<br />

59994 programs for different conditions.


4.E.A.2. Open a previously saved file: Press the Open<br />

button to open an existing file. Select the file you want to<br />

open and press OK.<br />

4.E.A.3. Copy a file: You can copy an existing file using<br />

copy function. The copied file allows you to further finetune<br />

each parameter if the growth conditions are similar.<br />

4.E.A.4. Rename a file: Rename an existing file using<br />

rename function.<br />

4.E.A.5. Delete a file: Delete saved files from the system<br />

memory.<br />

4.E.A.6. Display a file name:<br />

4.F. Acid-start setup<br />

This function is primarily used for Lactobacillus fermentation. To use this<br />

function, the user must input a pH value for Acid-Start setup, and then push<br />

the on the control page. Instead of immediately starting the<br />

automatic process, the controller will wait for the pH in the media to drop to<br />

the desired level, and then begin fermentation.<br />

63


4.G. pH-Auto start point<br />

Once the automatic process starts, the pH will be automatically controlled<br />

within this set value.<br />

4.H. Admin<br />

In this area, press the Admin button to enter account management dialog<br />

The supervisor has access to managing all the accounts. The supervisor can<br />

also change his/her or other users passwords through this page. Users have<br />

limited access to the controller. The functions that the user can use are<br />

assigned by the supervisor. The supervisor can set up a password for the<br />

remote control account.<br />

64


8.5 Chart<br />

65


66<br />

5.A.A.1. During the fermentation run, you can record the<br />

trend of the process. To record your process, go to the<br />

chart page.<br />

5.A.A.2. Press and start recording the process. The<br />

data will be saved into the built-in memory in log file<br />

format. The file will be named based on the date and time<br />

during which you record the process<br />

(Year/Month/Day/Hour/Min).<br />

5.A.A.3. If you want to stop recording a fermentation<br />

process, press the stop button to finish the recording.<br />

5.A.A.4. To read a saved file, press and select the xls<br />

file.<br />

5.A.A.5. Press to clear the chart screen<br />

5.A.A.6. Each parameter will be represented by different<br />

colors in the chart (Pump in yellow; DO in purple; pH in<br />

white; Agitation in green; Temp. in red)


5.A.A.7. You can change the scale for each parameter in<br />

the setting.<br />

67


-Sample rate: Recording frequency. For example, if the<br />

sample rate is 10 seconds, the system will collect data<br />

points every 10 seconds.<br />

-Temp.: Set the temp. display range<br />

-Rpm: Set the agitation speed display range<br />

-pH: Set the pH display range<br />

-PO2: Set the dissolved oxygen display range<br />

-ORP: Set the oxidation-reduction potential display range<br />

-CO2: Set the carbon dioxide display range<br />

-O2: Set the oxygen display range<br />

In order to help the user monitor the trend of a fermentation, we have the<br />

Long Curve and Short Curve displays. The purpose is to zoom in and<br />

zoom out of the curve. The data points of a short curve are fewer than those<br />

of a long curve.<br />

(6) X Scroll Gap: Scrolls the increments 20 units forward or backward every<br />

button.<br />

(8) X Marker: 50 stands for the X-axis coordinate displays. It goes in<br />

increments of 50 such as 0, 50, 100, 150, and so on.<br />

Data can be exported to an external disk for further analysis using the USB<br />

ports.<br />

Press Save to USB to export the files.<br />

68<br />

8.6 Calibration


6.A. DO probe calibration<br />

6.A.A.1. Press and enter the DO calibration page.<br />

6.A.A.2. Disconnect the cable from the DO probe. Press<br />

for Zero calibration.<br />

6.A.A.3. Place the probe into the vessel and connect the<br />

cable. Start the air pump and agitation to saturate the<br />

media dissolved oxygen.<br />

6.A.A.4. Press to set the DO 100% point when the<br />

sensor is fully polarized.<br />

6.A.A.5. Press to complete DO calibration.<br />

69


6.B. pH probe calibration<br />

pH calibration should be performed before sterilization.<br />

70


6.B.A.1. Press and start pH calibration<br />

6.B.A.2. Place the pH probe into the pH 7 standard solution.<br />

Wait until the pH value becomes stable, then press<br />

for Zero calibration.<br />

6.B.A.3. Clean the pH electrode with pure water, then dry it<br />

with Kimwipe tissue.<br />

6.B.A.4. Place the pH probe into the pH 4 standard solution<br />

and wait until the pH value is stabilized, and then press<br />

for Span calibration.<br />

6.B.A.5. Press to complete pH calibration.<br />

6.B.A.6. Clean the electrode again with pure water and dry<br />

it afterward. Disconnect the probe and install the probe<br />

into the vessel.<br />

6.C. Touch screen calibration: Use this function to align the<br />

screen position<br />

6.C.A.1. Press the Touch Calibration button. A green<br />

screen will appear.<br />

6.C.A.2. A flashing cross will appear in the corner for the<br />

first calibration point. Press and hold it down until the<br />

circle fills blue.<br />

6.C.A.3. Repeat with the remaining 3 points<br />

6.D. System time: Change the Year/Month/Hour/Min<br />

settings and press “ Set” to correct the system time<br />

71


8.7 Maint.<br />

This page is for mainly for manufacturer maintenance purposes and users<br />

have limited access.<br />

73


74<br />

7.A.A.1. Antifoam Detected Value: Foam level detection<br />

sensitivity can be increased/decreased by changing this<br />

value. This value has been optimized in the factory. There<br />

is no need to change the value. Default setting is 1800.<br />

7.A.A.2. Antifoam pulse interval: Settings for antifoam<br />

addition duration. T1 stands for pump on interval and T2<br />

stands for total duration.<br />

7.A.A.3. Oxygen pulse interval: If your system comes with<br />

an oxygen enrichment device(Model no.: FS-O-OE), you<br />

can set up this parameter. T1 stands for oxygen valve<br />

pulse cycle, T2 stands for on time.<br />

7.A.A.4. pH PID control: Set up the deadband (pH control<br />

range) for a pH control process. For example, if you want<br />

to control the pH between 7.2 and 6.8, set the deadband<br />

value at 0.2.


7.A.A.5. TCP/IP: Remote control internet settings (Please<br />

go to Section 10 for detailed settings)<br />

7.A.A.6. System logs: help users trace system profiles. The<br />

system log records a power failure, calibration raw data or<br />

error messages. Also, the system log records the user<br />

login time and info.<br />

-The calibration A/D value can be checked from calibration<br />

parameters dialog<br />

75


7.A.A.7. System Setup: Option to designate your controller<br />

according to your vessel model and the accessories that<br />

come with it. Choose motor 1 (M1 105 Watt) for vessels<br />

under 7 Liters, and choose motor 2 (M2 180 Watt) for 10,<br />

15 and 20 Liters. Choose DO probe type according to the<br />

probe you receive. Check the [CO2/O2 Gas Analyzer] box<br />

if you have the optional gas analyzer installed. System<br />

will automatically restart when the settings are applied.<br />

Note: After selecting and changing System Setup, the system will reboot<br />

automatically once you hit the Apply button<br />

Note: The PID value has been optimized before shipping. There is no need to<br />

change the PID settings.<br />

76


7.A.A.8. Restore cali.: This function allows you restore the<br />

factory settings. If you accidentally mess up your<br />

calibration profile, you can press this button to restore the<br />

factory settings.<br />

77


8.8 Help<br />

The help screen guides the user set up the cooling and air system after the<br />

vessel sterilization<br />

Section 9<br />

78


Section 1 Operating the system<br />

Note:<br />

79


80<br />

1.A.A.1. Cut lengths of tubing according to the distance<br />

between the connections.<br />

1.A.A.2. Bolded terms referring to the fermentation system<br />

can be found on the diagrams of the fermentation system<br />

(see Section 6 and Section 7).<br />

1.1 Preparation and connection of the control unit<br />

1.B. Examine the workplace<br />

1.B.A.1. Set up the control unit at the operating site and<br />

connect it to the required laboratory equipment:<br />

Power supply: make sure that the rated voltage is correct<br />

before plugging in the power cord.<br />

-Set up the coolant source: if you are using the chiller as the<br />

coolant source, please make sure the rated voltage is<br />

correct as well.<br />

1.B.A.2. See Section 6 for more details<br />

1.C. Set up the water circulation system<br />

1.C.A.1. Connect the coolant source (Chiller out port) to<br />

the Main water in port on the cooling system.<br />

1.C.A.2. Connect the coolant return (Chiller return port) to<br />

the Main water out (drain) port.<br />

1.C.A.3. Connect the cooling coil inlet to the Water in port<br />

and cooling coil outlet to the water out port.


1.C.A.4. Connect the condenser in to the condenser inlet<br />

and condenser out to the condenser outlet. (Please<br />

refer to the 8.8 Help and the picture on 7.2 Cooling<br />

system)<br />

81


9.2 Prepare the Culture Vessel and Accessories<br />

2.A. Check the Equipment<br />

While installing the culture vessel and fitting the probes and accessories,<br />

carefully check all parts for damage. Pay special attention to any potential<br />

cracks in the glass vessel, split in the O-rings (especially the one on the<br />

bottom of the head-plate), and any possible harm done to the seals on the<br />

rotating shaft.<br />

82<br />

Warning:


2.A.A.1. When a damaged glass vessel no longer has the<br />

bursting strength required for sterilization in the autoclave<br />

or running fermentation, it must be replaced immediately.<br />

2.A.A.2. Damage to the seal can cause contamination of the<br />

culture during the fermentation process. Seals-on<br />

accessories which are often removed and refitted should<br />

be replaced regularly.<br />

2.B. Vessel assembly and head-plate arrangements<br />

2.B.A.1. Unscrew the screws on top of the head-plate and<br />

remove the head-plate.<br />

2.B.A.2. Attach all parts (not including tubing) meant to be<br />

mounted onto the head-plate:<br />

2.B.A.3. O-ring seal: make sure the O-ring is properly fitted<br />

onto the bottom of the headplate<br />

2.B.A.4. Any port where there are no accessories part<br />

attached must be fitted with a blind port to prevent<br />

leakage.<br />

2.B.A.5. If an O-ring is damaged, it must be replaced. You<br />

may grease the O-rings (especially the one between<br />

glass vessel and headplate) with silicon grease to prevent<br />

the O-ring from deforming over time and to stop air from<br />

escaping from the vessel.<br />

2.C. Set up the air supply<br />

2.C.A.1. Attach two pieces of silicon tubing to either side of<br />

a 0.2 micron air filter (provided). Make sure the tubing<br />

lengths are long enough to reach from the controller to<br />

the vessel. Use clamps to close the tubing on each side<br />

of the filter.<br />

83


84<br />

2.C.A.2. Secure one end onto the To air sparger port on the<br />

vessel and the other onto the Gas out port on the<br />

controller<br />

2.C.A.3. With a storage bottle, connect one port with silicon<br />

tubing to the exhaust port on the top of the condenser.<br />

2.C.A.4. Connect the other port to a filter (an exhaust filter)<br />

with a short piece of silicon tubing.<br />

2.D. Set up the feeding port<br />

2.D.A.1. Fill the 250 ml storage bottles with acid, base,<br />

antifoam reagent, or nutrient.<br />

2.D.A.2. Connect them each with silicon tubing to the<br />

feeding ports on the head-plate. The tubing should be of<br />

sufficient length to allow for passage through the<br />

peristaltic pump in the control unit.<br />

2.D.A.3. If no acid/base/antifoam/nutrient is used, then use<br />

a shorter piece of silicon tubing to connect the two<br />

feeding ports in order to prevent air leakage.<br />

2.E. Media preparation<br />

2.E.A.1. Attach a piece of silicon tubing to the sampling port<br />

and clamp the other end.<br />

2.E.A.2. Adjust the height of the antifoam probe according<br />

the projected working volume. It is recommended that the<br />

user use at least 50 mm of tubing.<br />

2.E.A.3. Prepare the media and reagents.<br />

2.E.A.4. Pour the now mixed and ready-to-use media and<br />

water into the vessel.


2.F. pH calibration<br />

2.F.A.1. Connect the pH probe to the pH probe port on the<br />

controller.<br />

2.F.A.2. On the screen, select the Calibration tab and then<br />

select pH.<br />

2.F.A.3. Stick the probe into a pH 7 standard solution and<br />

wait until stable.<br />

2.F.A.4. Select Zero.<br />

2.F.A.5. Clean and dry the probe.<br />

2.F.A.6. Place the probe in pH 4 standard solution and wait<br />

until stable.<br />

2.F.A.7. Select Span Set.<br />

2.F.A.8. Press Finish then clean and dry the electrode<br />

again. (For more details in Section 8.6 calibration)<br />

2.G. DO calibration<br />

2.G.A.1. Select DO under the calibration tab.<br />

2.G.A.2. Make sure the DO probe and the DO probe port on<br />

the controller are not connected and then select Zero.<br />

2.G.A.3. Place the probe into the vessel and connect the<br />

cable.<br />

2.G.A.4. Start the air pump and agitation.<br />

2.G.A.5. Press Span Set when the sensor is fully polarized.<br />

85


86<br />

2.G.A.6. Press Finish. (For more details on 8.6 calibration)<br />

2.H. Final details<br />

2.H.A.1. Check all ports for proper connections.<br />

2.H.A.2. Secure the head-plate back onto the vessel by<br />

tightening the four or six screws.<br />

2.H.A.3. Make sure the o-ring on the bottom of the headplate<br />

is fitted well.<br />

2.H.A.4. Perform air leakage tests. If leakage happens,<br />

check again the possible leakage site.<br />

2.H.A.5. Wrap any exposed electrode part of the sensor<br />

(DO and pH) with aluminum foil to protect against the<br />

impact of steam.<br />

2.I.Sterilization of the culture vessel<br />

2.I.A.1. Take care that all open ports on the head-plate of<br />

the culture vessel are closed.<br />

2.I.A.2. Double-check that the electrode plugs are covered<br />

with aluminum foil.<br />

2.I.A.3. Completely disconnect the water circulation<br />

system.<br />

2.I.A.4. LOOSEN THE EXHAUST BOTTLE CAPS<br />

BEFORE STERILIZATION (Half-turn).<br />

2.I.A.5. Put the culture vessel with all storage and 500 ml<br />

exhaust bottle into the autoclave.


2.I.A.6. For single wall vessels, proper sterilization requires<br />

a temperature of 121°C for 40 minutes. For jacket<br />

vessels, the time required for sterilization is 60 minutes.<br />

9.3 Prepare the Fermentation Run<br />

3.A. Post-sterilization<br />

3.A.A.1. After sterilization, wear heat protective gloves and<br />

fasten the caps for all bottles immediately.<br />

3.A.A.2. After culture vessel and all bottles cool down, bring<br />

them from the autoclave to your work-place. Take caution<br />

during transportation when moving things from the<br />

autoclave to prevent the loosening of any tubing or<br />

connected assemblies.<br />

3.A.A.3. Place a piece of silicon tubing over the Air out port<br />

on the controller and insert the other end through a filter<br />

and then connect to the To air sparger port on the vessel.<br />

3.A.A.4. Connect the water in port on the controller to the<br />

cooling water in (coolant inlet) port on the vessel.<br />

3.A.A.5. Connect the water out port on the controller to the<br />

cooling water out (coolant outlet) port on the vessel.<br />

3.A.A.6. Connect the condenser in port on the controller to<br />

the condenser in port on the condenser.<br />

3.A.A.7. Connect the condenser out port on the controller to<br />

the condenser in port on the condenser (For more details<br />

on 8.6 DO calibration).<br />

3.A.A.8. Remove the protective aluminum foil from the<br />

electrode connectors.<br />

87


3.A.A.9. Connect the cable to the electrodes, including the<br />

pH probe, the pO2 probe and the foam level probe.<br />

3.A.A.10. Fill up the thermowell with water. Insert the Pt-100<br />

temperature probe.<br />

3.A.A.11. Connect the tubing from the feeding bottle to the<br />

feeding port.<br />

3.A.A.12. Place the silicon tubing through the peristaltic<br />

pump.<br />

3.B. Priming<br />

Perform priming to remove any air from the tubing:<br />

3.B.A.1. Select the Pump tab on the controller.<br />

3.B.A.2. Set the arrow to forward direction.<br />

3.B.A.3. Enter the desired rate at revolutions per minute.<br />

3.B.A.4. Press manual pumping button to perform priming<br />

(For more details on the manually adjusting the peristaltic pump options,<br />

please see section 8.2 Pump)<br />

88


3.C. Mount the motor<br />

3.C.A.1. Fit the bumper onto the stirring shaft.<br />

3.C.A.2. Locate the small notch on the motor and on the<br />

stirring shaft and place the motor onto the stirring shaft.<br />

3.C.A.3. Secure the motor by matching the latch.<br />

3.D. Cool down<br />

3.D.A.1. Make sure the condenser regulator is open<br />

halfway.<br />

3.D.A.2. Start the temperature PID control to the cool down<br />

the system (When you press the Temp. control button<br />

from the control interface, you will hear the sound of the<br />

solenoid valve. After cooling down the vessel to the<br />

working temp., wrap the heating blanket onto the culture<br />

vessel.<br />

DO NOT WRAP THE BLANKET AROUND THE VESSEL BEFORE<br />

IT HAS COOLED FROM STERILIZATION! This could result in the<br />

blanket damage (Blanket limitation: 60°C). Damage caused by<br />

mishandling will not be covered by our warranty.<br />

89


3.E. Inoculation<br />

3.E.A.1. Use silicon tubing to connect the inoculation bottle<br />

to the peristaltic pump (similar to the feeding bottle).<br />

3.E.A.2. Sterilize the feeding port and the silicon tubing. The<br />

user may apply a high temperature flame (1,200°C)<br />

toward the feeding port and silicon tubing opening for 10<br />

seconds.<br />

3.E.A.3. After proper sterilization of the feeding port,<br />

connect the silicon tubing to the feeding port.<br />

3.E.A.4. <strong>Manual</strong>ly pump the seed cells into the culture<br />

vessel. You can change the pump speed in the setup<br />

menu to achieve a higher flow rate.<br />

Note:<br />

Bore x<br />

wall<br />

90<br />

3.E.A.5. Tubing setup for the water-circulating system:<br />

Please go to the Help page to view the detailed<br />

connections.<br />

3.E.A.6. Half-turn the condenser water regulator to provide<br />

a continuous flow to the condenser<br />

3.E.A.7. Peristaltic pump<br />

-Quick and simple tube change.<br />

-Tubing is available in four sizes and five materials as<br />

follows:<br />

Autoprene Silicone Tygon Viton Prothane<br />

II


0.8mm x<br />

1.6mm<br />

1.6mm x<br />

1.6mm<br />

3.2mm x<br />

1.6mm<br />

4.8mm x<br />

1.6mm<br />

T0016 SAT 24 T0016 SLT 25<br />

T0016 SAT 07 T0016 SLT 12 T0016 TYG<br />

12<br />

T0016 SAT 08 T0016 SLT 16 T0016 TYG<br />

13<br />

T0016 SAT 01 T0016 SLT 02 T0016 TYG<br />

10<br />

T0016 VIT<br />

16<br />

T0016 VIT<br />

10<br />

T0016 VIT<br />

02<br />

T0016 PRO<br />

16<br />

T0016 PRO<br />

10<br />

T0016 PRO<br />

We provide #16 silicone tubing (3.2mm x 1.6mm) as the standard<br />

accessory. The silicone tubing is only for general purpose.<br />

Please do not use it for strong acid or base. To pump strong acid<br />

or base, please choose chemical resistant material (Viton or<br />

Tygon). Damage caused by misuse of liquids will not be covered<br />

by our warranty.<br />

91<br />

03


Section 10 Remote control setting<br />

Remote control is achievable using web browser, which is platform free.<br />

Additional installation of software is not required.<br />

10.1 Connecting your PC to the <strong>Winpact</strong> controller<br />

There are two ways to connect your PC to <strong>Winpact</strong> controller: direct<br />

connection from PC to controller, or connecting through a network<br />

1.A. Direct Connection (Controller to PC)<br />

Go to your PC, change your network settings so your network IP matches the<br />

controller IP.<br />

92<br />

1.A.A.1. Open your NETWORK Connection Setting window


1.A.A.2. Right click on Local Area Connection, and select<br />

Properties.<br />

93


94<br />

1.A.A.3. Choose TCP/IPv4 (this would differ depending on<br />

your operating system), and select Properties.


1.A.A.4. Change the IP to an IP thats in the same IP range<br />

shown on the controller. For example: default IP on the<br />

controller should be 10.0.0.166, change the IP on your PC<br />

to 10.0.0.55.<br />

1.B. Indirect Connection (Controller to Network)<br />

To connect through internet (LAN), you will have to setup the IP on the<br />

controller.<br />

Obtain a set of IP address that is within your network.<br />

Make sure that the IP you set for the controller is a FIXED IP, and that it does<br />

not conflict with any other IP that is within your network.<br />

95


Make sure to reboot the controller after changing the internet settings.<br />

No additional software is required, just make sure that JRE (Java Runtime<br />

Environment) is installed on your operating system.<br />

10.2 Controlling the <strong>Winpact</strong> system on your PC<br />

User: admin<br />

2.A.A.1. Open your web browser and key in the IP address<br />

with Port number. (Default IP is 10.0.0.166, Port: 8080)<br />

2.A.A.2. Click on Remote Control and the system will enter<br />

the login dialog. Type in the user name and password.<br />

Password: 0000 (Default password)<br />

96


2.A.A.3. Please skip the security dialog so that the login<br />

process can continue.<br />

97


2.B. Program page<br />

Monitor: The current value can be monitored from this page. The schematic<br />

diagram shows the working condition of each device. If the device is on, the<br />

color will be changed to red.<br />

98


2.C. Chart: Real-time recoding or history data reading can<br />

be done in this page. The log file can be saved into the<br />

computer disk.<br />

99


100<br />

2.D. <strong>Manual</strong> settings: In this page, you are able to perform<br />

remote control of agitation speed, temperature and pH.<br />

Press SET button to set up the parameters.


2.E. Auto settings: In this page, you can set up the<br />

schedule for the automatic fermentation process.<br />

Stirring table<br />

Temperature table<br />

101


Feeding table<br />

For security reasons, our remote control function will be disabled if a firewall<br />

exists. The remote control works fine within an intranet. If you have any<br />

problems with the internet setting, please consult your IT department.<br />

102


2.F. Exporting trend files into the USB hard drive for further<br />

analysis<br />

2.F.A.1. In the chart page, press Load File and select the<br />

file you wish to analyze.<br />

2.F.A.2. Plug the USB hard drive into USB port. Wait until<br />

the USB signal is detected, then press save to USB<br />

button to export the data.<br />

2.F.A.3. On your operating system, open the file to be<br />

analyzed.<br />

2.F.A.4. To plot the trend of a fermentation process, select<br />

the data you wish to plot. In this example, we select Time<br />

and Temp. for analysis.<br />

2.F.A.5. Use scatter plot function to plot the fermentation<br />

process (X axis: time, Y axis: Temp.).<br />

2.G. Understanding your data table<br />

103


Example (boxed in red, 2007/1/12, 01:02)<br />

Note:<br />

-Temperature: 48.7°C<br />

-Agitation speed: 299 rpm<br />

-pH: 6.61<br />

-pO2: 1%<br />

-Pump1: The pump has run for 51 seconds between<br />

2007/1/12 00:57 and 2007/1/12 01:02<br />

-Pump2 to Pump5: Off<br />

-Feed1 Total: the accumulated quantity of feeding before<br />

2007/1/12 01:02<br />

Parameter Unit<br />

Temperature Celsius ( ℃ )<br />

RPM Revolution Per Minute (RPM)<br />

pH pH<br />

PO2 Percentage (%)<br />

Pump 1 to Pump 5 Seconds (s)<br />

Feed 1 Total to Feed 5 Total Seconds (s)<br />

ORP millivolts (mV)<br />

104


CO2 Parts per million (PPM)<br />

O2 Percentage (%)<br />

105


Section 11 Cleaning and Maintenance<br />

11.1 Vessel Cleaning<br />

1.A.A.1. Fill the vessel with a mild detergent and water<br />

solution<br />

1.A.A.2. Let the solution stand for 1 hour, and then scrub<br />

thoroughly with a soft brush. A toothbrush may be used<br />

for smaller parts and hard to reach areas<br />

1.A.A.3. Drain, and rinse several times with tap water<br />

1.A.A.4. Rinse a final time with distilled water, and allow to<br />

dry<br />

11.2 Control Station Cleaning<br />

The control station can simply be cleaned by wiping with a damp cloth or<br />

towel<br />

11.3 DO electrode information and maintenance<br />

Please follow the instructions below to maintain your D.O. Probe:<br />

3.A. Cleaning the probe<br />

One effective way to remove contamination from the cathode surface is to<br />

wash it with a toothbrush and toothpaste. Then rinse with D.I. Water.<br />

3.B. D.O. sensor malfunction<br />

10 probable causes of D.O. sensor malfunction are listed below:<br />

106


3.B.A.1. Punctured membrane<br />

3.B.A.2. Torn or ripple membrane<br />

3.B.A.3. Dirty cathode<br />

3.B.A.4. Dirty or fouled anode surface<br />

3.B.A.5. Damaged O-ring<br />

3.B.A.6. Missing O-ring<br />

3.B.A.7. Damaged cathode<br />

3.B.A.8. Dirty membrane<br />

3.B.A.9. Corroded connector<br />

3.B.A.10. Electrolyte level too low<br />

Reference: Summer 1998 Broadley-James corporation<br />

3.C. Testing the D.O. Sensor membrane<br />

One simple way to test the membrane puncture is to connect the membrane<br />

cartridge with a syringe using an adapter. Monitor the bubble formation during<br />

pumping (See figure). Bubble formation may indicate leakage of the<br />

membrane.<br />

107


3.D. Membrane replacement and electrolyte refill<br />

3.D.A.1. Keep the sensor in an upright orientation, then<br />

unscrew the old membrane cap.<br />

3.D.A.2. Remove the membrane, expose the glass body in<br />

the air.<br />

3.D.A.3. Clean the tip of the glass body with a soft tissue.<br />

3.D.A.4. Gently introduce 1.5 ml DO solution into the new<br />

membrane cap.<br />

3.D.A.5. Carefully screw the new membrane cap onto the<br />

sensor shaft.<br />

How to replace the DO solution<br />

108


3.E. Membrane kit<br />

3.E.A.1. DO solution, 20ml<br />

3.E.A.2. Spare O-rings<br />

3.E.A.3. Pipette<br />

3.E.A.4. Polishing strip<br />

3.E.A.5. Membrane cartridge<br />

3.F. Storage and electrolyte replacement<br />

The DO probe can be stored in 3M Potassium Chloride solution. The DO<br />

solution is used for inner cartridge electrolyte replacement. After using the<br />

probe a few times, you can check the solution condition by nude eye (check if<br />

there is precipitation or color change). If precipitation or color change occurs,<br />

please replace with new solution.<br />

11.4 pH probe information and maintenance<br />

Please follow the instructions below for your pH electrodes maintenance:<br />

109


4.A. Cleaning the measuring electrode<br />

4.A.A.1. Remove any deposits from the membrane or<br />

diaphragm by rinsing the electrode with mild detergents.<br />

4.A.A.2. For calcium deposit, soak the electrode in a 0.1M<br />

HCl solution for a few minutes.<br />

4.A.A.3. For proteins, soak the electrode in a solution of 1%<br />

pepsin and 0.1M HCl for several hours.<br />

4.A.A.4. Inorganic coatings can be removed using<br />

commercially available glass cleaning solution (e.g.,<br />

Windex).<br />

4.A.A.5. Silver sulphide deposits, caused by the reaction of<br />

sulphides containing solution with silver chlorides in most<br />

electrodes, can be removed by soaking the electrode in a<br />

0.1 M thiourea/HCl solution until the diaphragm is totally<br />

bleached.<br />

4.A.A.6. To remove highly resistant deposits, hydrogen<br />

peroxide or sodium hypochlorite is recommended.<br />

4.A.A.7. Rinsing the electrodes in a 0.1M HCl or 0.1M<br />

NaOH solution for a few minutes may be able to remove<br />

the acid or alkaline soluble deposits.<br />

4.B. After cleaning<br />

After the cleaning process, soak the TrupH electrode in the storage solution<br />

(3M KCl) for 12 to 24 hours. Cleaning solution may penetrate the diaphragm<br />

during cleaning. It is necessary to calibrate the pH measurement after the<br />

hydration process.<br />

110


4.C. Storage of the electrode<br />

For long-term storage, immerse the pH electrode in a 3M KCl solution.<br />

4.D. Aging<br />

Aging is unavoidable. However, proper maintenance can delay the process.<br />

Typical symptoms of an aged measuring electrode includes:<br />

4.D.A.1. Increased response time<br />

4.D.A.2. Increased membrane resistance<br />

4.D.A.3. Declining slope, especially in the alkaline region<br />

4.D.A.4. Shift of the asymmetry potential (zero point shift)<br />

Lifetime of each electrode depends on the operating and handling conditions.<br />

The electrodes have a maximum operating lifespan of 18 months if handled<br />

properly. The lifetime may only last 2 months if operated at the temperatures<br />

exceeding 90°C. The electrode stops working after only 2 weeks when<br />

exposed to strong base (pH > 13) and high temperature (> 90°C).<br />

In most cases, we will find that pH electrodes age gradually. An increased<br />

response time is the most direct indication of aging. If the response time<br />

becomes unacceptable for your fermentation process control, we recommend<br />

replacing the pH electrode with a new one.<br />

4.E. Storage solution<br />

To prevent dehydration of the probe, we recommend storing the probe in the<br />

3M Potassium Chloride solution every time after use.<br />

111

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!