TROUBLESHOOTING<br />

The shutter will not open! Try putting the chamber door on again – the round button on<br />

The ‘No waterflow’ light is<br />

lit, and the machine<br />

appears to be off.<br />

The shutter is closing<br />

automatically of its own<br />

accord!<br />

the side of the X-ray tube housing must be fully pressed in<br />

be<strong>for</strong>e the instrument will recognise that it is safe to unleash<br />

the X-rays.<br />

If this button is fully in, check all the housing doors are fully<br />

closed and all interlocks in contact.<br />

There is also an interlock underneath the ‘top hat’<br />

arrangement (just on the chamber side be<strong>for</strong>e the secondary<br />

beam monochromator) – try pinching that closed.<br />

Finally, if the problem still persists, it is likely that one or<br />

both of the ‘Shutter Open’ bulbs has blown. Contact the Lab<br />

Manager.<br />

Last revised by N. Reeves-McLaren, 5 h January 2012.<br />

Low water pressure. Report to the Lab Manager.<br />

Probably someone has pressed the Shutter 2 ‘Clock’ button.<br />

Press the ‘∞‘ button, and try again.