10.5. Depth measurement<br />

10.5.1 Depth measurement (z-axis) with the<br />

combination drive is done by reading the scale<br />

on the drive knob during fine focussing between<br />

the two or more certain object levels. The scale<br />

is divided into 100 units. When the fine drive is<br />

used One graduation mark equals 2,8 µm. The<br />

range of depth measurement with combination<br />

drive is 0,288 mm.<br />

To avoid measuring errors caused by the<br />

reversal error of the drive (combination and<br />

coaxial coarse and fine drive) just measure the<br />

object points (object levels) from only one<br />

direction.<br />

10.5.2 Depth measurement with the coaxial<br />

coarse and fine drive is almost the same as with<br />

the combination drive. The scale is divided into<br />

50 units. When the fine drive is used one<br />

graduation mark equals 2,0 µm. The range of<br />

depth measurement with coaxial is limited to<br />

adjustable range of coarse drive (20 mm).<br />

10.5.2 Depth measurement can also be done<br />

with help of a analogue or digital dial indicator<br />

gauge which is mounted on the side. The gauge<br />

sensor cone is connected to the stage.<br />

pic. 23: indicator gauge for depth measurement<br />