MEASUREMENT POINT<br />
REGISTRATION<br />
Set the sensors so that they are<br />
approximately at the same rotational<br />
angle at the first measurement position.<br />
Select the register icon and<br />
press OK.<br />
This registers the first<br />
reading.<br />
Rotate the shafts to the next position.<br />
The shafts must be rotated over a<br />
minimum of 45°.<br />
Green sector show permitted positions.<br />
Red sector show forbidden positions.<br />
The Register icon is not shown if the<br />
rotation is less than 45°.<br />
<strong>Fixturlaser</strong> GO Pro Manual 3 rd edition 6.15