Digital Torque Transducer Data Sheet
Digital Torque Transducer Data Sheet
Digital Torque Transducer Data Sheet
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Bolted connection of the rotor<br />
Rotor<br />
Fastening bolts<br />
Bolt distribution view A<br />
B<br />
A<br />
Hexagon socket head bolts<br />
DIN EN ISO 4762; black/oiled/m tot =0,125<br />
(turned into the plane of projection)<br />
Bolt distribution view B<br />
Nominal (rated)<br />
torque<br />
(NVm)<br />
500<br />
1k<br />
2k<br />
3k<br />
Fastening bolts<br />
Property class of<br />
fastening bolts<br />
Prescribed<br />
tightening moment<br />
(NVm)<br />
M10<br />
10.9<br />
67<br />
M12<br />
115<br />
135<br />
5k M14 12.9<br />
220<br />
10k<br />
M16<br />
340<br />
Mounting dimensions<br />
Center of Rotor<br />
Center of Stator<br />
Measuring<br />
range<br />
500 N⋅m<br />
1 kN⋅m<br />
2 kN⋅m<br />
3 kN⋅m<br />
(Tolerance 1 mm)<br />
Mounting dimensions<br />
Mounting dimensions (mm)<br />
a b c<br />
2 2 0<br />
5 3 1<br />
5 kN⋅m 25 3 11<br />
10 kN⋅m 33 3 15<br />
Reserved additional space for fielbus cable:<br />
approx. 140 mm, from male connector area<br />
B1941-4.0 en<br />
a<br />
c<br />
b<br />
12<br />
HBM