Safety precautions for sensor handling<br />

Warning<br />

When the sensor is damaged or otherwise manipulated,<br />

alkaline (caustic) electrolyte may escape. With a pH<br />

value of 13, the electrolyte is strongly alkaline. Always<br />

avoid contact with skin, particularly eyes and mucous<br />

membranes.<br />

In case of a contamination immediately flush the affected<br />

parts with plenty of water.<br />

If you don’t feel well, immediately consult a doctor.<br />

When replacing the electrolyte and the membrane module,<br />

it is inevitable that electrolyte comes into contact with<br />

your fingers. Therefore be sure to wear protective gloves.<br />

Observe the technical specifications (see page 36)!<br />

<strong>Co</strong>mmissioning the SE 302 / SE 303 sensors<br />

Prior to initial<br />

start-up<br />

<strong>Co</strong>nnection<br />

Caution<br />

• It is recommended to replace the electrolyte as<br />

described on page 30 to compensate for electrolyte loss /<br />

consumption during storage.<br />

• Visually check the membrane for cracks or notches<br />

(When the membrane is damaged, the membrane<br />

module must be replaced, see page 30).<br />

• <strong>Co</strong>nnect the sensor to the meter observing the corresponding<br />

instructions (see page 14).<br />

Prior to initial start-up or after the sensor has been disconnected<br />

from the meter for more than 5 minutes, the sensor<br />

must be polarized before calibration or measurement.<br />

Sensors<br />

28