Safety precautions for sensor handling<br />
Warning<br />
When the sensor is damaged or otherwise manipulated,<br />
alkaline (caustic) electrolyte may escape. With a pH<br />
value of 13, the electrolyte is strongly alkaline. Always<br />
avoid contact with skin, particularly eyes and mucous<br />
membranes.<br />
In case of a contamination immediately flush the affected<br />
parts with plenty of water.<br />
If you don’t feel well, immediately consult a doctor.<br />
When replacing the electrolyte and the membrane module,<br />
it is inevitable that electrolyte comes into contact with<br />
your fingers. Therefore be sure to wear protective gloves.<br />
Observe the technical specifications (see page 36)!<br />
<strong>Co</strong>mmissioning the SE 302 / SE 303 sensors<br />
Prior to initial<br />
start-up<br />
<strong>Co</strong>nnection<br />
Caution<br />
• It is recommended to replace the electrolyte as<br />
described on page 30 to compensate for electrolyte loss /<br />
consumption during storage.<br />
• Visually check the membrane for cracks or notches<br />
(When the membrane is damaged, the membrane<br />
module must be replaced, see page 30).<br />
• <strong>Co</strong>nnect the sensor to the meter observing the corresponding<br />
instructions (see page 14).<br />
Prior to initial start-up or after the sensor has been disconnected<br />
from the meter for more than 5 minutes, the sensor<br />
must be polarized before calibration or measurement.<br />
Sensors<br />
28