Specifications for Model <strong>911</strong> <strong>Oxy</strong><br />

DO input<br />

Input for SE 302 / SE 303 dissolved oxygen sensors,<br />

various process sensors with 4-pole or VP cap<br />

Saturation 0 to 600 % (0 °C to 55 °C) 1)<br />

configurable to 0.0 to 199.9 % in the range 0 to 200 %<br />

Meas. range<br />

Measuring current 0 to 600 nA<br />

Meas. error 2,3,4) 0.5 % meas.val. + 0.2 %<br />

<strong>Co</strong>ncentration<br />

Meas. error 2,3,4)<br />

Barometric pressure<br />

Salt correction<br />

Calibration<br />

cal<br />

0.00 to 60.0 mg/l (0°C to 55°C)<br />

configurable to 0.00 to 19.99 mg/l in the range 0.0 to 20.0 mg/l<br />

0.5% meas.val. + 0.05mg/l<br />

700 to 1080 hPa<br />

<strong>Co</strong>nfigurable to 0.00 to 45.00 g/kg<br />

Solubility function to EN 25814<br />

0 = switched off, SA icon inactive<br />

Automatic with water-saturated air in quiver,<br />

Zero calibration possible<br />

Calibration range Zero point ±2 nA<br />

Polarization voltage<br />

Slope<br />

675 mV<br />

25 to 130 nA<br />

Cal timer 0 to 14 days default setting 6 days<br />

Pressure correction<br />

Temperature input<br />

Automatic using integrated pressure sensor<br />

NTC 22 kΩ integrated in DO sensor<br />

Meas. range -5.0 to +55.0 °C / 23 to +131 °F<br />

Resolution 0.1 °C / 1 °F<br />

Meas. error 2,3,4)<br />

Temperature<br />

compensation<br />

< 0.5 K<br />

Automatic with NTC<br />