<strong>prEN</strong> <strong>14931</strong>:2006 (E)<br />

Testing:<br />

Checking of whether the requirements according to IEC 60364-7-710 for spaces of the respective application<br />

category are complied with.<br />

4.2.22<br />

Chamber paint shall be compatible with hyperbaric conditions.<br />

4.3 Main chamber requirements<br />

4.3.1<br />

The maximum compression rate shall be greater than 80 kPa/min (0,8 bar/min) and not exceed 300 kPa/min<br />

(3,0 bar/min).<br />

Testing:<br />

Checking as to whether the values are complied with.<br />

4.3.2<br />

During normal operation it shall be possible for the pressure in the main chamber to be reduced from 40 kPa (0,4 bar)<br />

to 20 kPa (0,2 bar) within 1 min ± 20 s. Slower decompression rate during operation shall be possible. The main<br />

chamber shall be provided with a separate emergency discharge valve marked accordingly and sealed and permitting<br />

the pressure in the main chamber to be reduced from 200 kPa (2 bar) to ambient pressure within a period not<br />

exceeding 2 min.<br />

The location of the emergency discharge shall be defined using risk management procedure in accordance with<br />

EN ISO 14971 manufacturer in cooperation with the operator.<br />

Testing:<br />

Checking as to whether the values stipulated are complied with.<br />

4.3.3<br />

The number of occupants allowed in the main chamber is to be stated on a plate prominently and permanently fixed<br />

above the entrance.<br />

Testing:<br />

Inspection.<br />

4.3.4<br />

For each person to be accommodated, a seat or stretcher shall be provided and, after deduction for interior<br />

equipment, a volume within the chamber of at least 1 m³.<br />

Testing:<br />

Checking of number of seats by inspection; checking of space within the chamber on the basis of design documents.<br />

4.3.5<br />

The main chamber shall be equipped with means ensuring a ventilation to maintain the chamber atmosphere within<br />

specified limits given in Table 2. Means shall be provided so that during ventilation the pressure does not vary by<br />

more than ±5 %.<br />

12