<strong>prEN</strong> <strong>14931</strong>:2006 (E)<br />

⎯<br />

⎯<br />

⎯<br />

⎯<br />

⎯<br />

⎯<br />

⎯<br />

inside temperature for each compartment;<br />

oxygen concentration for each compartment;<br />

CO 2 partial pressure in the main chamber;<br />

supply pressure of the breathing system;<br />

status of the fire-extinguishing system (filling level, supply pressure) for each chamber;<br />

time;<br />

status of the electrical power supply.<br />

Testing:<br />

a) Inspection of controls and indicating instruments;<br />

b) checking as to whether controls and instruments are clearly marked and illuminated.<br />

4.5.2<br />

The pressure in the ante chamber and main chamber shall be indicated by at least one analogue pressure gauge<br />

each.<br />

Analogue pressure gauges shall have a minimum diameter of 120 mm and conform to at least precision class 0,25<br />

(see EN 837-1).<br />

Testing:<br />

Manufacturers' proof.<br />

Each main chamber shall be provided with a pressure test connection arranged such as to be easily accessible.<br />

Testing:<br />

Inspection.<br />

4.5.3<br />

Facilities shall be provided for continuously recording the pressure in each ante chamber and main chamber. The<br />

system shall indicate pressure variations of 3 kPa (0,03 bar) and time intervals of 1,0 min in a manner which enables<br />

the data to be evaluated. The pressure variations over the previous 3 h shall be visible.<br />

Testing:<br />

Inspection to check whether equipment is provided for recording the working pressure in the ante chamber and main<br />

chamber.<br />

Manufacturers' proof regarding the pressure gauge of the recording instruments.<br />

4.5.4<br />

Continuous measurement, display and recording of the oxygen concentration in percent by volume shall be possible<br />

separately for each ante chamber and main chamber. The sampling point shall be located close to the ventilation<br />

outlet. For chambers designed to accommodate more than 12 persons a second measuring device shall be installed.<br />

This second sampling point should be located close to the point where the highest oxygen concentration is to be<br />

expected.<br />

16