MAN-MA35-e
MAN-MA35-e
MAN-MA35-e
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Step<br />
Key (or instruction) Display<br />
The prompt for calibration P ˚oC<br />
weight is displayed<br />
min<br />
A<br />
E<br />
% MS<br />
g /l<br />
CAL SET PRG END START<br />
min<br />
TAR<br />
8. Open the hinged cover<br />
9. Place the 30-weight on P ˚oC<br />
the weighing system<br />
Minus sign –: weight value too low<br />
Plus sign +: weight value too high<br />
No sign: weight value OK<br />
The weight unit symbol (g) is P ˚oC<br />
displayed at the end of adjustment<br />
min<br />
A<br />
E<br />
% MS<br />
g /l<br />
CAL SET PRG END START<br />
min<br />
A<br />
E<br />
% MS<br />
g /l<br />
CAL SET PRG END START<br />
min<br />
TAR<br />
min<br />
TAR<br />
Printout after calibration and --------------------<br />
adjustment 23.08.2005 10:51 *)<br />
*) Date and time included only Model <strong>MA35</strong>-000230V1<br />
if a Sartorius printer model Ser. no. 99992581<br />
YDP02 or YDP03 is used Ver. no. 00-33-01<br />
ID<br />
--------------------<br />
External calibration<br />
W-ID<br />
Nom. + 30.000 g<br />
Diff. + 0.001 g<br />
External adjustment<br />
Diff. + 0.000 g<br />
completed<br />
--------------------<br />
Name:<br />
10. Unload the analyzer<br />
Close the cover<br />
11. Quit calibration/adjustment c<br />
--------------------<br />
25