Step<br />

Key (or instruction) Display<br />

The prompt for calibration P ˚oC<br />

weight is displayed<br />

min<br />

A<br />

E<br />

% MS<br />

g /l<br />

CAL SET PRG END START<br />

min<br />

TAR<br />

8. Open the hinged cover<br />

9. Place the 30-weight on P ˚oC<br />

the weighing system<br />

Minus sign –: weight value too low<br />

Plus sign +: weight value too high<br />

No sign: weight value OK<br />

The weight unit symbol (g) is P ˚oC<br />

displayed at the end of adjustment<br />

min<br />

A<br />

E<br />

% MS<br />

g /l<br />

CAL SET PRG END START<br />

min<br />

A<br />

E<br />

% MS<br />

g /l<br />

CAL SET PRG END START<br />

min<br />

TAR<br />

min<br />

TAR<br />

Printout after calibration and --------------------<br />

adjustment 23.08.2005 10:51 *)<br />

*) Date and time included only Model <strong>MA35</strong>-000230V1<br />

if a Sartorius printer model Ser. no. 99992581<br />

YDP02 or YDP03 is used Ver. no. 00-33-01<br />

ID<br />

--------------------<br />

External calibration<br />

W-ID<br />

Nom. + 30.000 g<br />

Diff. + 0.001 g<br />

External adjustment<br />

Diff. + 0.000 g<br />

completed<br />

--------------------<br />

Name:<br />

10. Unload the analyzer<br />

Close the cover<br />

11. Quit calibration/adjustment c<br />

--------------------<br />

25