HYPROP manual - UMS
HYPROP manual - UMS
HYPROP manual - UMS
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Get ready to start a measuring campaign<br />
Manually Refilling (incl. delivery):<br />
Alternatively, take the reservoir syringe, fill it with completely<br />
degassed water 1 and take care to avoid bubbles in front of the<br />
ceramic. Fill the shaft with water and plug the vacuum syringe filled<br />
with ¼ deionized degased water completely over the thread. Pull it,<br />
until both snappers are locked. Now, the water from the reservoir<br />
syringe is drawn through the ceramic tip into the vacuum syringe.<br />
When approximately 10 ml are flushed through the ceramic the<br />
<strong>HYPROP</strong> shaft is filled.<br />
Figure 4-7<br />
1 Push out all air from the syringe. Now plug the end of the tube with your finger and<br />
pull up the syringe. This creates vacuum inside the syringe and dissolved gas is<br />
released. Rotate the still evacuated syringe to collect all bubbles from the wall of the<br />
syringe. Hold the syringe upright and slide in the piston. Unblock the tube and push<br />
out all air. Repeat this procedure until no bubbles are produced anymore.<br />
26/104