Get ready to start a measuring campaign<br />

Manually Refilling (incl. delivery):<br />

Alternatively, take the reservoir syringe, fill it with completely<br />

degassed water 1 and take care to avoid bubbles in front of the<br />

ceramic. Fill the shaft with water and plug the vacuum syringe filled<br />

with ¼ deionized degased water completely over the thread. Pull it,<br />

until both snappers are locked. Now, the water from the reservoir<br />

syringe is drawn through the ceramic tip into the vacuum syringe.<br />

When approximately 10 ml are flushed through the ceramic the<br />

<strong>HYPROP</strong> shaft is filled.<br />

Figure 4-7<br />

1 Push out all air from the syringe. Now plug the end of the tube with your finger and<br />

pull up the syringe. This creates vacuum inside the syringe and dissolved gas is<br />

released. Rotate the still evacuated syringe to collect all bubbles from the wall of the<br />

syringe. Hold the syringe upright and slide in the piston. Unblock the tube and push<br />

out all air. Repeat this procedure until no bubbles are produced anymore.<br />

26/104