03.11.2014 Views

INSTRUmENTS - Dentsply

INSTRUmENTS - Dentsply

INSTRUmENTS - Dentsply

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Small things. Big difference.<br />

STERILISATION<br />

burs from damage. Burs left standing wet may stain or corrode.<br />

Diamond/tungsten burs: Burs should be reprocessed as soon as is<br />

reasonably practical following use.<br />

PREPARATION FOR CLEANING: No specific requirements.<br />

Disassembly not required.<br />

AUTOMATIC CLEANING: Equipment: Use only validated washing<br />

machines (HTM 2030: UK only) and validated cleaning agents<br />

suitable for both devices and washing machine. Follow instructions<br />

issued by both the equipment and cleaning agent manufacturers.<br />

When unloading, inspect for complete removal of visible soil.<br />

If necessary, repeat cycle, or use manual cleaning before<br />

repeating cycle.<br />

MAINTENANCE: Discard damaged burs.<br />

INSPECTION AND ALL BURS: Visually inspect to ensure all<br />

contamination has been removed. Check for damage and wear.<br />

Cutting Function Testing: edges should be free of defects. Discard<br />

damaged or corroded burs. Check burs with long, slender features<br />

for distortion. Where burs form part of a larger assembly, check<br />

assembly with mating components. There may be evidence of<br />

discolouration or dulling after sterilisation. This is normal and will not<br />

affect performance.<br />

DRYING: Dry using paper towelling or dry heat not<br />

exceeding 140°C.<br />

PACKAGING FOR STERILISATION: The UK Dept. of Health<br />

recommends that validated, pre-vacuum autoclaves are used for<br />

sterilising wrapped burs Sterilisation: Autoclaves drawing only a<br />

post-sterilisation vacuum for drying purposes are not suitable for<br />

processing wrapped burs. Burs may be loaded into dedicated<br />

instrument trays or general-purpose sterilisation trays. Ensure that<br />

cutting edges are protected.<br />

STERILISATION: Use only validated vacuum autoclaves (Refer to<br />

HTM2010: UK only). Do not exceed 140°C.<br />

Sterilisation Temperature Range (°C) Nominal Pressure (MPa) Minimum Time at Temperature* (mins)<br />

Minimum<br />

Maximum<br />

134 137 0.225 3<br />

* Minimum holding time when nominal operating temperature and pressure have been reached.<br />

STORAGE: Avoid contamination.<br />

ADDITIONAL INFORMATION: When sterilising multiple burs in one<br />

autoclave cycle, ensure that the manufacturers stated maximum load<br />

is Information: not exceeded. Validated ultrasonic baths may be used<br />

with validated cleaning agents. Follow instructions for use issued by<br />

both the equipment and cleaning agent manufacturers.<br />

MANUFACTURER CONTACT: See brochure for telephone and<br />

address of local representative or telephone DENTSPLY Ash<br />

Instruments on +44 (0)1932 853422.<br />

normally determined by wear and damage due to use. Do not use<br />

chemical disinfectants prior to sterilisation or rapid deterioration of the<br />

material may occur. Cold liquid disinfection/sterilisation, chemical<br />

vapour sterilisation, and dry heat sterilisation methods have not been<br />

tested or validated for efficacy and are not recommended for use.<br />

INSTRUCTIONS<br />

POINT OF USE: Remove excess soil with disposable cloth or paper.<br />

CONTAINMENT AND TRANSPORTATION: Protect instruments from<br />

damage. Instruments should be reprocessed as soon as is reasonably<br />

practical following use. Instruments left standing wet may stain or<br />

corrode.<br />

CLEANING - Automated: May be cleaned using an ultrasonic water<br />

bath for 15 minutes.<br />

CLEANING: Manual: Rinse all instruments under running water to<br />

remove any gross debris.<br />

DISINFECTION: Additional disinfection is not necessary prior<br />

to sterilisation.<br />

PACKAGING: Place instrument in a paper or paper/plastic steam<br />

sterilisation pouch. Instruments may be loaded into dedicated<br />

instrument trays.<br />

STERILISATION: Use a steam autoclave. Place bagged instruments<br />

into the steam autoclave, paper side up when using a paper/<br />

plastic pouch. After warm-up is completed, operate at a sterilizing<br />

temperature and pressure of 273ºF/134ºC to 279ºF/137ºC and<br />

pressure of 31psi/216kPa for 6 to 12 minutes. Alternate method:<br />

Place non-bagged inserts into the steam autoclave. After warm-up<br />

is completed, operate at a sterilizing temperature and pressure of<br />

273ºF/134ºC to 279ºF/137ºC and pressure of 31psi/216kPa for<br />

3 to 4 minutes. Instrument should be used immediately.<br />

DRYING: To dry, use the drying cycle of the autoclave. Set cycle for<br />

20-30 minutes.<br />

MAINTENANCE, INSPECTION AND TESTING: Visually inspect<br />

to ensure all contamination is removed. Visually inspect all o-rings<br />

for wear. Replace worn o-rings (Refer to O-Ring Replacement Kit<br />

Instructions). Lubricate o-rings with water before use. Check for<br />

distortion, damage and wear (Refer to Efficiency Wear Indicator<br />

Instructions). Discard damaged, worn or corroded instruments.<br />

STORAGE: To maintain sterility, instruments should remain<br />

bagged until ready for use.<br />

CAVITRON INSERTS: INFECTION<br />

CONTROL PROCEDURES<br />

WARNINGS<br />

Instruments are not sterile upon receipt and must be sterilized prior to<br />

use in accordance with the following instructions. The use of cleaning<br />

and disinfectant solutions is not recommended.<br />

LIMITATIONS ON PROCESSING: Repeated cleaning and sterilisation<br />

cycles have minimum effect on these instruments. End of life is<br />

EARN REWARDS dentsplyrewards.co.uk<br />

CONTACT US dentsply.co.uk 0800 0723313<br />

LEARN ABOUT IT dentsplyacademy.co.uk<br />

207

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!