07.11.2014 Views

Gaymar FW600 Operator's Manual

Gaymar FW600 Operator's Manual

Gaymar FW600 Operator's Manual

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

ENGLISH<br />

REMOVE FROM USE<br />

indicator<br />

STANDBY<br />

indicator<br />

SETPOINT switch<br />

button<br />

IV pole clamp<br />

RUN /STANDBY<br />

switch<br />

Cassette slot<br />

Bubble trap<br />

receiver<br />

Figure 1—<strong>FW600</strong> series Blood/Fluid Warmer<br />

2.0 Product Description<br />

The Medi-Temp III <strong>FW600</strong> Series Blood/Fluid Warmer<br />

is a dry-heat device designed for safe and rapid<br />

warming of blood, blood products and other fluids<br />

through the utilization of disposable blood/fluid<br />

warming sets. The warmer does not provide fluid flow<br />

rate control. Blood, blood products and other fluids<br />

normally refrigerated, can be rapidly warmed to user<br />

selectable temperatures of between 38.0°C–43.0°C<br />

at flow rates up to 300 ml/min. Room temperature<br />

fluids can be warmed at flow rates up to 500 ml/min.<br />

The disposable Blood/Fluid warming sets are available<br />

in Standard, Standard with Extensions, Pediatric, and<br />

High Flow models.<br />

2.1 Warmer<br />

The warming unit is designed to be IV pole<br />

mounted. The device has no on/off switch.<br />

Connection to the AC supply is indicated by a<br />

green STANDBY mode indicator, located on the<br />

front panel. In STANDBY mode, no power can be<br />

applied to the heater. To begin normal operation,<br />

the RUN/STANDBY button must be pressed.<br />

When this occurs, an LED temperature display located<br />

on the front panel will display the actual fluid<br />

temperature or the setpoint temperature in °C.<br />

The setpoint is user selectable from 38.0°C–43.0°C<br />

in 1.0°C increments, by use of the SETPOINT switch on<br />

the front panel. Also located on the front of the unit is<br />

a slot for insertion of the disposable warming set. On<br />

the right side of the unit, just below the IV pole clamp,<br />

is a receiver for the disposable warming set’s bubble<br />

trap.<br />

2.2 Disposable sets<br />

The Medi-Temp III <strong>FW600</strong> Series Blood/Fluid warming<br />

unit is designed to be used only with <strong>Gaymar</strong> D25000<br />

series disposable warming sets. The following warming<br />

sets (figure 2) are designed for use up to 300 mm Hg<br />

pressure. Contact your local dealer for more<br />

information.<br />