Operator’s <strong>Manual</strong><br />
Calibration<br />
The amplifier is calibrated at the factory prior to being shipped. The recommended calibration<br />
interval is one year. Calibration should be performed by qualified personnel only. Schedule an<br />
annual factory calibration as part of your regular maintenance. Extended warranty, calibration, and<br />
upgrade plans are available for purchase. Contact your <strong>Teledyne</strong> <strong>LeCroy</strong> sales representative or<br />
customersupport@teledynelecroy.com to purchase a service plan.<br />
Power<br />
AC Power Source<br />
100 to 240 VAC (±10%) at 50/60 Hz (± 10%).<br />
<strong>Manual</strong> voltage selection is not required because the instrument automatically adapts to line<br />
voltage.<br />
Power Consumption<br />
<strong>DA</strong><strong>1855A</strong>: 28 Watts (39 VA)<br />
<strong>DA</strong><strong>1855A</strong>-PR2: 56 Watts (78 VA)<br />
Power and Ground Connections<br />
The amplifier is provided with a 10A/250V 18AWG rated grounded cord set containing a molded<br />
three-terminal polarized plug and a standard IEC320 (Type C13) connector for making line voltage<br />
and safety ground connections.<br />
The AC inlet ground is connected directly to the frame of the instrument. For adequate protection<br />
again electric shock, connect to a mating outlet with a safety ground contact.<br />
WARNING. Interrupting the protective conductor inside or outside the device, or<br />
disconnecting the safety ground terminal, creates a hazardous situation. Intentional<br />
interruption is prohibited.<br />
Power On/Off<br />
The Power On/Off switch on the back of the amplifier controls the operational state of the device.<br />
Press the side of the switch closest to either On or Off.<br />