TREATMENT PLANNING SYSTEM<br />

The Radiotherapy treatment planning system should be fully computerised, integrated<br />

system having hardware and software to perform all kinds of 3D Teletherapy and Brachytherapy<br />

planning calculations, isodose plotting and display of patient files, beam data acquisition and<br />

display and other related programme.<br />

1. HARDWARE:<br />

(a) Work Station: -<br />

The Treatment planning system should have two separate computers<br />

(in addition to the control of the HDR Brachytherapy machine) and should<br />

have a most modern graphics workstation with all validated & advanced<br />

features.<br />

(b)Film scanner/ Digitizer with Light Box:-<br />

The system should have a film scanner preferably or a back lighting digitizer<br />

with digitizer cursor for transferring the information from X ray, patient contour,<br />

CT scan films, and images from CT-SIM etc into the system.<br />

(c) Printer/Plotter:-<br />

The system should have a fast multi-colour plotter preferably A/3 size to print<br />

out various datas and Isodose curves along with a good quality laser printer<br />

WORK STATION<br />

Hardware:<br />

• Latest available high speed graphic workstation and windows<br />

- Optical mouse and WIN keyboard<br />

• 100 GB hard disc or more<br />

• DVD Writer<br />

-Monitor - LED 20" Color, flat panel with high resolution<br />

-Software components:<br />

- Window 7<br />

- Norton Antivirus with 5 yrs validity and for more than 1 user.<br />

- 1000GB extra data storage device -5 in number<br />

- PC anywhere (only for the server)<br />

- Low noise version and long-time reliability due to temperature control<br />

SOFTWARE FEATURES TELETHERAPY<br />

Contouring:-<br />

- Volume definition should be possible using Volume Segmentation using threshold, Free hand<br />

contour tracing, Contour editing, 3D anistrophic Margins etc.<br />

- Volume delineation should be possible with Free hand contour tracking or Advanced volume<br />

segmentation using threshold in 3D or with predefined shapes. Various contour editing tools to<br />

modify the contour at any plane should be possible.<br />

- It is desirable to have the facility to contour in Axial, Sagittal, and Coronal or in any<br />

oblique planes.<br />

- It should be possible to do manual, semi-automated, fully automated contouring /<br />

segmentation in the images.<br />