<strong>Feeding</strong> Formula VTBD<br />
1-3000 mL in 1 mL increments<br />
Bolus Volume<br />
1-3000 mL in 1 mL increments<br />
Number of Boluses<br />
1-99<br />
Bolus Interval<br />
1-24 hours in 1-hour increments<br />
Flushing Solution Dose Range<br />
10-500 mL in 1 mL increments<br />
Flushing Solution Interval Range<br />
1-24 hr in 1 hr increments<br />
Accuracy<br />
±10% or 0.5 mL/hr, whichever is larger, with bag at 46 cm (18”) above pump, at room temperature 22° C ± 2° C (72° F ± 3° F), using<br />
a new pump set with less than the recommended 24 hours of maximum usage.<br />
Occlusion Pressure<br />
15 psi (103 kPa) Nominal<br />
Dimensions<br />
Height: 16.8 cm (6.6”) Width: 16.3 cm (6.4”) Depth: 11.7 cm (4.6”)<br />
Weight<br />
1.1 kg (2.4 lbs), 1.2 kg (2.7 lbs) with pole clamp<br />
Material<br />
Housing: ABS/Polycarbonate blend<br />
<strong>Pump</strong> Set Door: Polyester/ Polycarbonate blend<br />
Power<br />
Use AC adapter for wall outlet usage. <strong>Pump</strong> operates on 9V DC, 1.5 Amps.<br />
Battery<br />
New, fully charged NiMH Battery Pack delivers ≥ 15 hrs at 100 mL/hr feed rate. Approximately 15 minutes prior to complete battery<br />
discharge, a low battery alarm will occur (see subsection “Battery Low” in Section IX). When complete discharge occurs, the pump<br />
will automatically turn itself off.<br />
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required to fully<br />
recharge the battery pack.<br />
Alarms<br />
• System Error<br />
• Flow Error<br />
• <strong>Pump</strong> Set Dislodged Error<br />
• Hold Error<br />
• Rotor Error<br />