3-4<br />

<br />

The detector is used with a probe protector to support the probe (<strong>ZR22G</strong>) when the probe length<br />

is 2.5 m or more and it is mounted horizontally.<br />

(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in the<br />

probe insertion hole.<br />

(2) Make sure that the sensor assembly mounting screws (four bolts) at the tip of the detector are<br />

not loose.<br />

(3) Mount the detector so that the reference gas and calibration gas inlet faces downward.<br />

Flow direction of the<br />

sample gas<br />

1050,1550,2050<br />

Gasket (t1.5)<br />

Unit: mm<br />

Φ60.5<br />

Notch<br />

Probe top<br />

Reference gas inlet<br />

Calibration gas inlet<br />

Figure 3.3 Probe protector (supporting the mounting strength)<br />

F3-3E.ai<br />

<br />

The detector is used with a probe protector to prevent the sensor from being worn by dust<br />

particles when there is a high concentration of dust and gas flow exceeds 10 m/sec (pulverized<br />

coal boiler or fluidized-bed furnace).<br />

(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in the<br />

probe insertion hole. The probe protector should be installed so that the notch is downstream<br />

of the sample gas flow.<br />

(2) Make sure that the sensor assembly mounting screws (four bolts) at the probe tip are not<br />

loose.<br />

(3) Where the detector is mounted horizontally, the reference gas and calibration gas inlet should<br />

face downward.<br />

NOTE<br />

When the probe protector is used in the <strong>ZR22G</strong> with pressure compensation (-P), instrument air<br />

leaking from the probe protector may affect the measured value.<br />

1050,1550,2050<br />

Gasket (t1.5)<br />

Unit: mm<br />

Direction of the sample gas flow<br />

Probe top<br />

Mount the protector so that the notch is<br />

the downstream side of the gas flow.<br />

Figure 3.4 Mounting of detector with a probe protector<br />

F3-4E.ai<br />

Reference gas inlet<br />

Calibration gas inlet<br />

IM 11M12A01-02E<br />

8th Edition : Jan.13,2012-00