3-4<br />
<br />
The detector is used with a probe protector to support the probe (<strong>ZR22G</strong>) when the probe length<br />
is 2.5 m or more and it is mounted horizontally.<br />
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in the<br />
probe insertion hole.<br />
(2) Make sure that the sensor assembly mounting screws (four bolts) at the tip of the detector are<br />
not loose.<br />
(3) Mount the detector so that the reference gas and calibration gas inlet faces downward.<br />
Flow direction of the<br />
sample gas<br />
1050,1550,2050<br />
Gasket (t1.5)<br />
Unit: mm<br />
Φ60.5<br />
Notch<br />
Probe top<br />
Reference gas inlet<br />
Calibration gas inlet<br />
Figure 3.3 Probe protector (supporting the mounting strength)<br />
F3-3E.ai<br />
<br />
The detector is used with a probe protector to prevent the sensor from being worn by dust<br />
particles when there is a high concentration of dust and gas flow exceeds 10 m/sec (pulverized<br />
coal boiler or fluidized-bed furnace).<br />
(1) Put a gasket (provided by the user) between the flanges, and mount the probe protector in the<br />
probe insertion hole. The probe protector should be installed so that the notch is downstream<br />
of the sample gas flow.<br />
(2) Make sure that the sensor assembly mounting screws (four bolts) at the probe tip are not<br />
loose.<br />
(3) Where the detector is mounted horizontally, the reference gas and calibration gas inlet should<br />
face downward.<br />
NOTE<br />
When the probe protector is used in the <strong>ZR22G</strong> with pressure compensation (-P), instrument air<br />
leaking from the probe protector may affect the measured value.<br />
1050,1550,2050<br />
Gasket (t1.5)<br />
Unit: mm<br />
Direction of the sample gas flow<br />
Probe top<br />
Mount the protector so that the notch is<br />
the downstream side of the gas flow.<br />
Figure 3.4 Mounting of detector with a probe protector<br />
F3-4E.ai<br />
Reference gas inlet<br />
Calibration gas inlet<br />
IM 11M12A01-02E<br />
8th Edition : Jan.13,2012-00