02.01.2015 Views

AMMTIAC Quarterly, Vol. 2, No. 2 - Advanced Materials ...

AMMTIAC Quarterly, Vol. 2, No. 2 - Advanced Materials ...

AMMTIAC Quarterly, Vol. 2, No. 2 - Advanced Materials ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

System Expenses<br />

At approximately twice the cost, typical CR systems are more<br />

expensive to purchase than film processing systems. In general,<br />

depending upon system configuration, conventional CR<br />

systems approach a purchase cost of $125,000. It is important<br />

to note that these costs are dropping, while the cost of the<br />

film-based system is staying the same or increasing. Regarding<br />

recurring costs, each IP costs approximately $550 to $700,<br />

nearly equal the cost for 100 sheets of film. An important difference<br />

is that these plates can be reused up to 5000 times.<br />

The IPs require occasional cleaning and other maintenance.<br />

While climate control for the IP storage is not necessary, moisture<br />

can be a considerable problem. Moisture and the presence<br />

of dirt and grime will shorten the life of the IP.<br />

There are two pieces of support equipment used for traditional<br />

radiography methods that will still be needed in a CR<br />

system. These are film identification units and some type of<br />

storage cabinets for CD-ROMs.<br />

Figure 3. A Closer View of the Inspection Results.<br />

curve. Standards for accepting and rejecting inspection results<br />

are still being developed. Although many recurring expenses are<br />

reduced, some still exist and the up-front costs of CR are substantial.<br />

Finally, the complexity of CR systems warrants careful<br />

consideration on where they will be located by those planning<br />

to implement the systems.<br />

The Learning Curve<br />

Conventional film-based radiography has well established procedures<br />

for carrying out radiographic techniques. These procedures<br />

include the amount of radiation, length of exposure and<br />

resulting image quality. Because, as previously noted, CR typically<br />

requires less radiation and shorter exposure time, operators<br />

need to learn how to optimize the parameters in order to<br />

achieve acceptable results. In addition, the spatial resolution of<br />

the images (i.e., how coarse or fine the image is) affects the<br />

interpretation of the results. There is concern that until training<br />

and standards are well established and coordinated, images<br />

potentially could be over-analyzed, and anomalies that would<br />

have been acceptable under wet film processing may now cause<br />

the part being inspected to fail the test.<br />

Standards for Accepting or Rejecting Results<br />

Current accept/reject standards are based on film as well as<br />

the proven history of how defects will appear in a film-based<br />

system. Changing the capabilities of the imaging system also<br />

changes the predictability of results. Further research is needed<br />

to compile a catalog of digital results which are subject to<br />

an analytical process that addresses probability of detection<br />

(POD), probability of failure, desired or expected service life,<br />

etc. This type of work is still in the early stages of progress.<br />

Pending new standards, CR will not be accepted for certain<br />

types of inspections.<br />

System Location Requirements<br />

The intended installation location should be carefully planned<br />

and evaluated prior to proceeding with the implementation due<br />

to the complexity of CR systems. The temperature of the housing<br />

facility should be stable, and there should be no heat<br />

sources (including direct sunlight) within close proximity. As<br />

noted above, moisture and excessive dust and corrosive gases<br />

would also degrade system performance; humidity and ventilation,<br />

therefore, should be considered. Constant vibration and<br />

shock must be avoided as well.<br />

COMPUTED RADIOGRAPHY SYSTEMS IN NAVY REGION<br />

NORTHWEST<br />

CNRNW acquired CR systems (see Figure 4) through the<br />

PPEP for three separate Navy activities. Two activities received<br />

their systems in 2003 and have used them for several informational<br />

inspections with favorable results. Based on those results,<br />

CNRNW requested an additional system from PPEP for a<br />

third activity; it was installed in early 2006.<br />

Figure 4. An NDI Inspector Uses an IP Reader and Workstation.<br />

http://ammtiac.alionscience.com The <strong>AMMTIAC</strong> <strong>Quarterly</strong>, <strong>Vol</strong>ume 2, Number 2 13

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!