AMMTIAC Quarterly, Vol. 2, No. 2 - Advanced Materials ...
AMMTIAC Quarterly, Vol. 2, No. 2 - Advanced Materials ...
AMMTIAC Quarterly, Vol. 2, No. 2 - Advanced Materials ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
System Expenses<br />
At approximately twice the cost, typical CR systems are more<br />
expensive to purchase than film processing systems. In general,<br />
depending upon system configuration, conventional CR<br />
systems approach a purchase cost of $125,000. It is important<br />
to note that these costs are dropping, while the cost of the<br />
film-based system is staying the same or increasing. Regarding<br />
recurring costs, each IP costs approximately $550 to $700,<br />
nearly equal the cost for 100 sheets of film. An important difference<br />
is that these plates can be reused up to 5000 times.<br />
The IPs require occasional cleaning and other maintenance.<br />
While climate control for the IP storage is not necessary, moisture<br />
can be a considerable problem. Moisture and the presence<br />
of dirt and grime will shorten the life of the IP.<br />
There are two pieces of support equipment used for traditional<br />
radiography methods that will still be needed in a CR<br />
system. These are film identification units and some type of<br />
storage cabinets for CD-ROMs.<br />
Figure 3. A Closer View of the Inspection Results.<br />
curve. Standards for accepting and rejecting inspection results<br />
are still being developed. Although many recurring expenses are<br />
reduced, some still exist and the up-front costs of CR are substantial.<br />
Finally, the complexity of CR systems warrants careful<br />
consideration on where they will be located by those planning<br />
to implement the systems.<br />
The Learning Curve<br />
Conventional film-based radiography has well established procedures<br />
for carrying out radiographic techniques. These procedures<br />
include the amount of radiation, length of exposure and<br />
resulting image quality. Because, as previously noted, CR typically<br />
requires less radiation and shorter exposure time, operators<br />
need to learn how to optimize the parameters in order to<br />
achieve acceptable results. In addition, the spatial resolution of<br />
the images (i.e., how coarse or fine the image is) affects the<br />
interpretation of the results. There is concern that until training<br />
and standards are well established and coordinated, images<br />
potentially could be over-analyzed, and anomalies that would<br />
have been acceptable under wet film processing may now cause<br />
the part being inspected to fail the test.<br />
Standards for Accepting or Rejecting Results<br />
Current accept/reject standards are based on film as well as<br />
the proven history of how defects will appear in a film-based<br />
system. Changing the capabilities of the imaging system also<br />
changes the predictability of results. Further research is needed<br />
to compile a catalog of digital results which are subject to<br />
an analytical process that addresses probability of detection<br />
(POD), probability of failure, desired or expected service life,<br />
etc. This type of work is still in the early stages of progress.<br />
Pending new standards, CR will not be accepted for certain<br />
types of inspections.<br />
System Location Requirements<br />
The intended installation location should be carefully planned<br />
and evaluated prior to proceeding with the implementation due<br />
to the complexity of CR systems. The temperature of the housing<br />
facility should be stable, and there should be no heat<br />
sources (including direct sunlight) within close proximity. As<br />
noted above, moisture and excessive dust and corrosive gases<br />
would also degrade system performance; humidity and ventilation,<br />
therefore, should be considered. Constant vibration and<br />
shock must be avoided as well.<br />
COMPUTED RADIOGRAPHY SYSTEMS IN NAVY REGION<br />
NORTHWEST<br />
CNRNW acquired CR systems (see Figure 4) through the<br />
PPEP for three separate Navy activities. Two activities received<br />
their systems in 2003 and have used them for several informational<br />
inspections with favorable results. Based on those results,<br />
CNRNW requested an additional system from PPEP for a<br />
third activity; it was installed in early 2006.<br />
Figure 4. An NDI Inspector Uses an IP Reader and Workstation.<br />
http://ammtiac.alionscience.com The <strong>AMMTIAC</strong> <strong>Quarterly</strong>, <strong>Vol</strong>ume 2, Number 2 13