Chapter 3 - BREATH TYPES<br />

This chapter contains information regarding the breath types available on the <strong>LTV</strong> ® <strong>1200</strong> / 1150<br />

ventilator. It covers how breaths are initiated, limited and cycled, and when each type of breath is<br />

given.<br />

The following terms are used in discussing how breaths are given:<br />

Initiate<br />

Limit<br />

Cycle<br />

What causes a breath to be given. Breaths may be initiated by a patient trigger, a<br />

push of the <strong>Manual</strong> Breath button, or by the ventilator based on the set breath rate<br />

and ventilation mode.<br />

How the breath is controlled. Breaths may be limited to a maximum circuit pressure<br />

or flow.<br />

What causes the breath to be cycled from the inspiratory phase to the exhalation<br />

phase. Breaths may be cycled by the ventilator when a set time or delivered volume<br />

has been reached, or when an alarm condition such as a high pressure limit has<br />

been reached.<br />

Breath Types<br />

Breaths are defined by how they are initiated, limited and cycled. The breath types are Machine,<br />

Assist, and Patient.<br />

Machine Assist Patient 8<br />

Initiated By - Ventilator Patient Patient<br />

Limited By - Ventilator Ventilator Ventilator<br />

Cycled By - Ventilator Ventilator Patient<br />

Breaths may be given in any of the following forms: Volume Control, Pressure Control, Pressure<br />

Support and Spontaneous.These breaths are given as described in the sections below.<br />

In addition, the following parameters apply to all breaths:<br />

The Minimum Inspiratory Time is 300 ms.<br />

The Minimum Exhalation Time is 346 ms.<br />

When patient triggers are enabled, triggers are detected during exhalation after the Minimum<br />

Exhalation Time has expired.<br />

8 Pressure Support or Spontaneous<br />

Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> / 1150 Ventilators Page 3-1<br />