Measurement Functions<br />

THD+N<br />

In this mode the <strong>Minilyzer</strong> measures the THD+N (Total Harmonic<br />

Distortion + Noise) and the k2 - k5 harmonic distortion.<br />

Input Level<br />

(RMS)<br />

Fig 3.10 THD+N Meter Panel<br />

The measurement is carried out in the band from 10 Hz - 20 kHz,<br />

expressed in dB or in %. Simultaneously, the LCD shows the actual<br />

RMS input level or the 2nd, 3rd, 4th & 5th harmonics distortion below<br />

the THD+N result.<br />

NOTE<br />

The THD+N results are calculated using a restricted<br />

measuring bandwidth of 10 Hz - 20 kHz.<br />

Besides checking the linear purity of a sine signal - e.g. measuring<br />

amount of harmonic distortions - this measurement mode is<br />

particularly suited to get a quick idea whether unwanted disturbances<br />

like hum are present.<br />