Measurement Functions<br />

vu + PPM<br />

<strong>ML1</strong> provides vu + PPM (Peak Program Meter) measurement results,<br />

following IEC60268.<br />

Reference<br />

Level<br />

Numerical<br />

Peak-Hold<br />

Graphical<br />

Peak-Hold<br />

Integration Time<br />

Reference Indicator<br />

Fig 3.11 vu + PPM (Type I) Panel<br />

The PPM may be measured according to three standards:<br />

• Type I<br />

• Type IIA<br />

• Nordic standard<br />

One of these standards has to be selected when entering the vu+PPM<br />

mode through the measurement functions menu.<br />

The following chapters describe the different entry fields and indicators<br />

of the vu+PPM display.<br />

a. Reference Level & Indicator<br />

The reference level indicates the RMS level of a 1 kHz sine wave that<br />

defines the value of the reference indicator. The reference level may<br />

be defined by the user in a range of -20 dBu to +14 dBu.<br />

The default reference level depends on the selected standard<br />

• +4 dBu for the vu-meter<br />

• +6 dBu for the PPM standards Type I and Nordic<br />

• +8 dBu for the PPM standard Type IIA.<br />