The differential input mode can measure ±20 V signals in the ±20 V range.<br />

However, the maximum voltage on any one pin is ±10 V with respect to<br />

GND. For example, if AI 1 is +10 V and AI 5 is –10 V, then the<br />

measurement returned from the device is +20 V.<br />

20<br />

15<br />

10<br />

Amplitude (V)<br />

5<br />

0<br />

–5<br />

–10<br />

AI 1<br />

AI 5<br />

Result<br />

–15<br />

–20<br />

Figure 8. Example of a Differential 20 V Measurement<br />

Connecting a signal greater than ±10 V on either pin results in a clipped<br />

output.<br />

20<br />

15<br />

10<br />

Amplitude (V)<br />

5<br />

0<br />

–5<br />

–10<br />

AI 1<br />

AI 5<br />

Result<br />

–15<br />

–20<br />

Figure 9. Exceeding ±10 V on AI Returns Clipped Output<br />