20 <strong>Lumax</strong> <strong>Family</strong> Technical Manual<br />

Category<br />

Table 4: Summary <strong>of</strong> Complications—Tupos LV⁄ATx<br />

# <strong>of</strong><br />

Pts<br />

% <strong>of</strong><br />

Pts<br />

LV Lead Related<br />

# <strong>of</strong><br />

Complications<br />

Complications<br />

per pt‐year<br />

High threshold⁄<br />

Intermittent<br />

biventricular 11 8.27% 12 0.12<br />

capture⁄ No<br />

capture<br />

Unable to implant<br />

lead via coronary 11 8.27% 11 0.11<br />

sinus<br />

Dislodgement 4 3.01% 4 0.04<br />

Diaphragmatic⁄<br />

Intercostal<br />

1 0.75% 2 0.02<br />

stimulation<br />

Total 27 20.3% 29 0.28<br />

Device Related<br />

Infection 3 2.26% 7 0.07<br />

Device migration 4 3.01% 4 0.04<br />

Elective<br />

replacement 4 3.01% 4 0.04<br />

indicator reached<br />

Inductions <strong>and</strong><br />

conversions<br />

1 0.75% 1 0.01<br />

Unable to<br />

interrogate device<br />

1 0.75% 1 0.01<br />

Total 12 9.02% 17 0.17<br />

Total Procedure<br />

<strong>and</strong> Device<br />

Related<br />

43 32.33% 58 0.57