PAGE 20<br />

DIAGENODE MYCOBACTERIUM TUBERCULOSIS REAL-TIME PCR USER MANUAL<br />

Troubleshooting guide<br />

> Nucleic Acid isolation<br />

Refer to the manufacturer’s product insert<br />

NB:<br />

Perform preventive maintenance of workstations for automated extraction (according to the<br />

manufacturer’s recommendations).<br />

Check the expiration date and storage conditions for solutions provided<br />

> PCR<br />

The signal detected for the positive <strong>Mycobacterium</strong> tuberculosis control, the DNA extraction<br />

& inhibition control (DIA-EIC/DNA-050) are negative or weak.<br />

The PCR conditions do not comply with the protocol:<br />

1. Check extraction procedure.<br />

It’s important to select an appropriate extraction procedure. The extraction procedure<br />

must be validated and limit the presence of inhibitory factors.<br />

2. Check PCR reaction set-up.<br />

To obtain a high sensitivity, you must respect volumes described and the master mixes<br />

suggested.<br />

NB:<br />

Check the calibration of pipettes used (according to the manufacturer’s recommendations).<br />

Check the expiration date and storage conditions of master mixes suggested.<br />

3. Check the PCR profile.<br />

To obtain a high sensitivity, respect the protocol adapted to the real-time PCR used.<br />

4. Check dyes selected.<br />

Check and select dyes adapted for each real-time PCR platform.<br />

5. Interpretation of results.<br />

Check the table and select the appropriate channel.<br />

Important note:<br />

Check storage conditions (during and after experiments) and expiration date.These two<br />

conditions highly influence results obtained.<br />

> Real-time PCR system<br />

Refer to the manufacturer’s product insert.<br />

NB:<br />

Perform maintenance of the real-time PCR platform (according to the manufacturer’s<br />

recommendations).<br />

For trouble shootings/questions, please contact us:<br />

Tel.: +32 (0)4 364 20 50 - info@diagenodediagnostics.com<br />

Europe <strong>Diagenode</strong> sa / CHU - Tour GIGA - B34 - 3 e étage // Avenue de l’Hôpital, 1 // 4000 Liège (Sart Tilman) // Belgium // Phone: (+32) 4 364 20 50 // Mail: info@diagenode.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!