4 Accessory units 25<br />

4.1.4.4 Mounting of <strong>pr</strong>ism wash systems with high <strong>pr</strong>essure water<br />

!<br />

Warning! In high <strong>pr</strong>essure wash systems, <strong>pr</strong>essure increase can occur in a closed pipe section when the<br />

high <strong>pr</strong>essure pump is operated. K–<strong>Patents</strong> recommends to mount a <strong>pr</strong>essure relief valve in the pipe section.<br />

Relief <strong>pr</strong>essure should be according to pipe <strong>pr</strong>essure rating.<br />

7<br />

6<br />

5<br />

FLOW<br />

10<br />

2<br />

9<br />

0<br />

POWER RELAY UNIT<br />

PR-36<strong>03</strong>-___<br />

POWER<br />

MAIN<br />

SWITCH<br />

4<br />

3<br />

12 8<br />

13<br />

11<br />

TAP WATER, TEMP MAX 60 °C / 140 °F<br />

3/8"<br />

13 CABLE 2x1 (AWG 17) CUSTOMER 1<br />

12 11 POWER SUPPLY CABLE 3 METERS FOR HIGH PRESSURE PUMP POWER SUPPLY CABLE 3 METERS FOR WATER VALVE<br />

K-PATENTS CUSTOMER<br />

1<br />

10 HIGH PRESSURE PUMP PR-3602-110/230/400/550 INCLUDING WATER VALVE K-PATENTS<br />

9 POWER RELAY UNIT PR-36<strong>03</strong>-110/230/400/550 K-PATENTS 1<br />

8 AC POWER SUPPLY 230/110 V<br />

CUSTOMER<br />

7 HIGH PRESSURE HOSE 8M<br />

K-PATENTS<br />

6 CHECK VALVE R 1/4" PR-3302<br />

K-P / CUSTOMER<br />

5 EFC (ELBOW FLOW CELL)<br />

K-PATENTS<br />

4 CABLE BETWEEN INDICATING TRANSMITTER AND SENSOR PR-8300<br />

K-PATENTS<br />

3<br />

2<br />

1<br />

WASH RELAY UNIT (WR)<br />

SENSOR PR-<strong>03</strong>-A62-HSS<br />

INDICATING TRANSMITTER IT-R<br />

K-PATENTS<br />

K-PATENTS<br />

K-PATENTS<br />

PART PART SPECIFICATIONS<br />

SUPPLIED BY<br />

1<br />

1<br />

2<br />

1<br />

1<br />

1<br />

1<br />

1<br />

1<br />

1<br />

POWER<br />

1<br />

8<br />

I<br />

Figure 4.9<br />

Mounting summary of <strong>pr</strong>ism wash system <strong>for</strong> high <strong>pr</strong>essure water with -WR relay unit