kENDALL KANGAROO JOEY™
kENDALL KANGAROO JOEY™
kENDALL KANGAROO JOEY™
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Figure 31. Pump Set Dislodged Error screen.<br />
Battery Low (Fixed Audible Volume)<br />
The BATTERY LOW screen appears and the alarm beeps continuously when the battery needs to be recharged. There is<br />
approximately 15 minutes of battery life remaining when this screen appears, Figure 32.<br />
Plug the power cord in to begin recharging. The pump will automatically return to the screen that was active prior to the error. The<br />
battery will charge continuously whenever the pump is plugged into a wall outlet. 6 hours of charging is required to fully recharge<br />
the battery pack.<br />
If this screen appears while the A/C Adaptor is plugged in to the pump, check to make sure the A/C Adaptor plug is pushed all the<br />
way in so it is flush with the back of the pump.<br />
Figure 32. Battery Low error screen.<br />
Indicates that the battery must be<br />
recharged immediately.<br />
Feeding Complete (Adjustable Audible Volume)<br />
The FEEDING COMPLETE information screen appears after completion of the programmed feeding, Figure 33. Press4Power Down<br />
to turn off the pump. Press4CONTINUE to return to the SET LOADED opening menu, Figures 7A-7D.<br />
Figure 33. Feeding Complete<br />
notification screen.<br />
Set Use >24 Hours Warning<br />
The pump set usage warning indicator will blink on the RUNNING screen if a pump set has been used for 24 or more hours (hours<br />
actually running). It is recommended to replace pump sets after this length of usage. This icon is only an informational message and<br />
does not require action.<br />
Figure 34 shows an example of the RUNNING screen, where a Feed & Flush Set has been in use for more than 24 hours. The<br />
informational warning indicator shows up in the upper left on the screen.<br />
<strong>KANGAROO</strong> JOEY 30 Enteral Feeding Pump