II.6 Test ResultsII.6.1Test PassedWhen a yield test runs successfully, the resultant Yield Stress (Pa), the % Torque value at thetime of the yield and the Temperature at yield are printed.Test (#5: Sample Tst) CompleteYield Stress (Pa) = 726.36 % Torque @ Yield = 73.56 Temperature = 25.5CIn addition, the results are shown on the display as follows:II.6.2 Test Failed%=73.56 025.5CYIELD=726.36 PaFigure 21Some of the reasons a yield test may fail are as follows.II.6.2.1 Under-Range ConditionOnly yield measurements in the % Torque range from 10% to 100% will be accepted. Ifa Yield Stress is reached before the % Torque value is 10%, an Under-Range error causes thetest to end.The following is sent to an attached printer:Test (#5: Sample Tst) CompleteTest Failed = Under-rangeThe following screen is displayed:YIELD ERRORUNDER-RANGEFigure 22II.6.2.2 Over-Range ConditionOnly yield measurements in the % Torque range from 10% to 100% will be accepted. IfYield Stress is not reached before the measured % Torque value reaches 100%, an Over-Range error causes the test to end.The following is sent to an attached printer:Test (#5: Sample Tst) CompleteTest Failed = Over-rangeThe following screen is displayed:YIELD ERROROVER-RANGEFigure 23Brookfield Engineering Labs., Inc. Page 14 <strong>Manual</strong> No. M/02-215

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!