MXF-2400 - Shimadzu do Brasil Comércio

MXF-2400 - Shimadzu do Brasil Comércio

MXF-2400 - Shimadzu do Brasil Comércio

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

SpecificationsSpecifications1. X-ray Fluorescence SpectrometerElements to be determined: 4Be ~ 92UElements determined simultaneously: Up to 36 elementsMode of analysis: Simultaneous determination of many elementsAtmosphere: Vacuum, air, helium (optional)■Spectrometer UnitFixed monochrometer: Converging method with curved crystal.Vacuum type for all the elements.Crystal: SX, TAP, PET, Ge, NaCl, LiFDetector: Gas sealed detector for 11Na ~ 92UNe, Ar, Kr exatronNe, Ar, Kr multitronFPC for 5B, 6C, 8O, 9FScanning monochrometer (optional): Parallel beam method with flat crystal.Used for heavy elementsCrystal: LiFDetector: Scintillation counter (SC)Elements to be determined: 22Ti ~ 92U (with crystal for heavy element determination)Present mechanism: Up to 30 elements may be preset■Measuring ElectronicsType: Pulse counting system for all the elements controlled with a microprocessorCounting capacity: 4 x 10 9 counts/elementCounting method: integration for the preset timeIntegration time: Adjustable for each group, from 1 to 999 secondsHigh voltage supply for detector: 1,550 to 2,150 V■X-ray GeneratorX-ray tube: End win<strong>do</strong>w type with Rh target.X-ray win<strong>do</strong>w: Beryllium win<strong>do</strong>w (Thin Win<strong>do</strong>w)Maximum output: 4 kWX-ray power controllerRectification: Full wave rectification and smoothing with a capacitorControl method: Secondary side detection, primary side control, control with the CPUMaximum rating: 50 kV, 100 mA, 4 kWStability: ±0.01% (for 10% source fluctuation), both the current and voltageTube voltage setting: 5 kV steps from 20 to 50 kVTube current setting: 50 kA steps form 5 to 100 mASafety circuit: Against overvoltage, overcurrent, overload, and abnormal cooling watersupply■Sample Feeding MechanismType: Accurate sample positioning with a swing arm type sample feederTurntable: Accepts 8 samples togetherSpinner: 60 rpm (50/60 Hz)Sample holder: 64 mmf, 43 mm highMax. sample size: 51 mmf, 38 mm high■Evacuation SystemOil rotary pump:Evacuation rate: About 130 L / 160 L /min. (50/60 Hz)Oil mist filter providedVacuum gauge: Pirani gauge, constant temperature typeVacuum stabilizer: Controlled with a microcomputer. The degree of vacuum canbe stabilized at an arbitrary point.■Gas Supply System for Flow Proportional CounterUse: Used of determination of B, C, N, O, and FGas: PR gas (Ar 90% and CH4 10%)Gas flow consumption: 10 to 15 mL /min.Gas density stabilizer: Controlled by a microcomputerFlow control: Needle valve and flow meterGas cylinder: A 6 Nm 3 cylinder (If light elements (4Be~9F) are attached)■X-ray ProtectionWarning: Yellow lamp on the front panel and red lamp on the X-ray control panel.Safety device: If the sample setter and the X-ray shutter are opened at thesame time, the X-ray tube will be automatically turned off.Also the moment the panel is opened the X-ray output is cutoff.■X-ray Tube Cooling Unit (CWC-16, incorporated in the main body)Type: Circulation of distilled water (a 18-liter tank is attached.)Method of heat exchange: Carried out in a dual tube between externalcooling water and the distilled water.Heat exchange capacity: 4 kW (3.440 kcal / hour)Water purity maintenance: By ion exchangeAlarm: Temperature, flow rate, and electric conductivityExternal cooling water: Necessary (Not necessary if using HYCOOL 30)2. Data Processing Unit■Hardware (IBM PC / AT compatible)OS: Win<strong>do</strong>ws XPMain Memory: 256MB or moreKeyboard: Full keyboardFloppy disk: 3.5"<strong>do</strong>uble-sided, <strong>do</strong>uble density (1.44 MB / disk). Single drive.Hard disk: 10GB or morePrinter: Laser printer■SoftwareProgram for quantitative analysisMaximum number of elements to be processed: Arbitrary number of elementsMaximum number of elements to be simultaneously determined: 36 per sampleNumber of analysis groups: Arbitrary number of groupsRepeated analysis: Arbitrary number of timesPriority interrupt analysis: Possible, automatic return to sequential analysisPrintout: Printout order can be specified for each group. Compound names andelement names may be printed out. Pass and fail marks may be printed out.Number of calibration curves: Per each group, per each elementType of calibration curve: Second-order polynomialCorrection for dead timeCorrection for absorption and enhancement: Per each group, per each elementMaximum number of elements to be corrected for absorption and enhancement: Arbitrarynumber of elements per equationCorrection for drift: 2-point method or 1-point methodCorrection for overlapping: 2-point method or 1-point methodInternal standard method: Available for each group or each elementChemical correction: 2-point method or 1-point methodNumber of calculation formula: Arbitrary channels for concentration calculationProgram for scanner: Arbitrary number of elementsProgram for qualitative analysisScanner: For heavy element determinationMethod: Step scanningMethod of processing: Automatic peak search and automatic peak element determinationData output: 0n display and printerFactor calculationCalibration factor: By the method of least squares, up to second-order polynomialFactor for correcting absorption and enhancement: By the method of multiple regression(concentration correction and sensitivity correction)Maintenance programPulse height distribution: Graphic displayMonochromator alignment aid: Graphic displayCounter unit: Graphic displayAlarm: Display of alarm number, cause, and measure.Manual instrument diagnosis: Parts by part check is possible.External transmission program: Analysis data may be transmitted externally, in qualitativeanalysis.External transmission (analysis results, errors, etc.)LAN and RS-232C (optional)Automatic notification via e-mail possible20

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!