7.Enhanced Sensitivity for Superlight ElementsThe layered structure analyzer (LSA), which uses curved crystals for convergence, provides almost ten times highersensitivity for X-rays than conventional detectors.Unlike the conventional dispersing crystals, the LSA has synthetic multilayer structures.The spacer is made of the material and has the thickness that gives the best lattice constant to sense X-rayintensities at the highest sensitivity for such elements as Be, B, C, N, O, F, Na, and Mg.TAP. (Mg, Na, F)12LSA dispersing elementLSA dispersing elementX-ray intensity ratio111098765432Monochromator for LSA1Mg Ma F ON C B BeElementX-ray Intensity Comparison between LSA Dispersing Elementand TAP Crystal Total Reflection8.Wide Dynamic RangeThe <strong>Shimadzu</strong>’s original pulse counting circuit and the counting error correcting circuit combine to provide a widedynamic range: one calibration curve can cover the concentration range more than 3,000,000 cps, from 0% to 100%.High accuracy is ensured even in high concentration region where the X-ray intensity is very high.( x 10 4 CPS)300X-ray tube voltage: 40 kVX-ray:CuKαX-ray intensity2001000 10 20 3040 50 60Linear Calibration Curve more than 3,000,000 cpsX-ray tube current (mA)9