<strong>WEIGHT</strong> <strong>INDICATOR</strong>7 Equipment Faults7.1 Out of range<strong>PR430S2</strong>This condition occurs if the load cell input signal is outside the fullscale range as defined by the amplifier pre-gain (parameter PGunder 4.2.3 above).Display shows "ADC-SAT+"Indicates that the input is outside the range in the positive direction.Either the signal is too large due to a load cell fault or the pre-gain(PG) is set too high.Display shows "ADC SAT-"Indicates that the input is outside the range in the negative direction,i.e. below zero.Display shows "SENSE ER"Indicates that the sense voltage is no longer within 3V of the voltageregistered at calibrationIssue 3.3 December, 2010 Page 39 of 40