12.07.2015 Views

GPR-18 MS ATEX Explosion Proof PPB Oxygen Analyzer.

GPR-18 MS ATEX Explosion Proof PPB Oxygen Analyzer.

GPR-18 MS ATEX Explosion Proof PPB Oxygen Analyzer.

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Advanced Instruments Inc.Zero Calibration OffsetAdjustment Capability:Designed to facilitate precise analysis below 5% of the most sensitive or lowest rangeavailable on the analyzer, this feature enables users’ to compensate for background offsets,as defined above, of up to 50% of the most sensitive or lowest full scale range available onthe analyzer and bring analyzers online faster.As described below, accomplishing either objective places a degree of responsibility on theuser. Determining the true offset requires the user to wait, see Online Recovery Timesection, until the analyzer reading is no longer trending downward (best evidenced by aconstant horizontal trend on an external recording device). Bringing the analyzer onlinefaster, basically the same as choosing not to wait for the stable horizontal trend reading,requires the user to repeat the ZERO CALIBRATION function. The frequency of which is atthe user’s discretion, hourly is recommended but at least when the reading goes negative.Advanced Instruments’ oxygen analyzers are capable of zero offset adjustments of 50% ofthe most sensitive or lowest range available on the analyzer. Since every analyzer is QCtested to 1% of the most sensitive or lowest range available, exercise CAUTION wheninitiating the ZERO CALIBRATION function at 50% (prematurely) of the most sensitive orlowest range available on the analyzer. If the anticipated O2 analysis level is less than theoffset value or if adequate time is not allowed for the analyzer to establish the true offset,the analyzer will in all probability display a negative reading.Note: From the SYSTEM menu option “Display Negative (Reading)” users can togglebetween ON and OFF choose by pressing the ENTER key and control whetheranalyzer displays negative readings.Type of <strong>Analyzer</strong>:Online ppb or ppm process analyzers: Analysis below 5% of the most sensitive or lowestrange is normally limited to these analyzers. However, such analysis is possible withportable analyzers from Advanced Instruments due to their 60 day duty cycle and/or abilityto operate during battery charging.Portable analyzers: Typically used intermittently moving between different samplepoints/systems for trend analysis above 5% of the most sensitive range and, therefore, theyfall into the “all other analysis” category requiring only span calibration.Percentage analyzers: Generally used above 5% of the most sensitive range and, therefore,fall into the “all other analysis” category requiring only span calibration.Online Recovery Time:The priority users place on getting or keeping an analyzer online is “the” most significantfactor involved in calibration and troubleshooting issues. The time it takes an analyzer tocome down to a specific level after exposure to high O2 concentrations or air is significantlydifferent if a sensor is being installed than if the sensor had been in-service at low ppmlevels for more than 1 week as illustrated below:Sensor Calibration at Install In-service Calibration<strong>MS</strong> Pico-Ion Air to 1 ppm < 45 minAir to .1 ppm < 6 hrsAir to .02 ppm < 8 hrs80 ppm to .02 ppm < 10 min9 ppm to .02 ppm < 5 minppm Fuel Cell Air to 100 ppm < 10 minAir to 10 ppm < 1 hrAir to 1 ppm < 6 hrsAir to .1 ppm < 16 hrs% Fuel Cell Air to .1% < 1 minAir to .05% < 3 minAir to 1 ppm < 30 min80 ppm to .1 ppm < 10 min8 ppm to .05 ppm < 5 minThe above times assume the introduction of a purge gas, the lower of the available zero orsample (if known and constant) gas, with a ppm level O2 concentration less than 0.5 ppm isintroduced to the analyzer following span calibration to purge (accelerate the reaction of)the O2 that has dissolved into the electrolyte inside the sensor. If zero gas is not available,substitute the sample gas and expect slightly longer recovery time.23

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!