Functional Description<br />
The <strong>HESI</strong>-<strong>II</strong> probe produces charged aerosol droplets that contain sample ions. The <strong>HESI</strong>-<strong>II</strong><br />
probe accommodates liquid flows of 1 μL/min to 1 mL/min without splitting.<br />
The removable components of the <strong>HESI</strong>-<strong>II</strong> probe are the end cover that is secured to the<br />
probe body with socket head screws and the needle insert that screws into the probe body<br />
(see Figure 3). The end cover includes the high voltage feedthrough and the sample inlet port.<br />
The external components of the probe body include the grounding union holder, vaporizer<br />
cable connector socket, probe sleeve with depth markers, and probe nozzle (see Figure 3 and<br />
Figure 4).<br />
Figure 3. <strong>HESI</strong>-<strong>II</strong> probe removable components<br />
End cover<br />
Socket head screws<br />
Needle insert<br />
<strong>Probe</strong> body<br />
<strong>Thermo</strong> <strong>Scientific</strong> <strong>HESI</strong>-<strong>II</strong> <strong>Probe</strong> <strong>User</strong> <strong>Guide</strong> 5<br />
2<br />
<strong>Probe</strong> sleeve with<br />
depth markers