use a conventional gauge dead weight tester to calibrate the <strong>202BG</strong>, but only if thecommon-mode corrections are known beforehand.Conformance with CEC Standard<strong>202BG</strong> Deviation (Pa)3.002.251.500.750.00-0.75-1.50-2.25-3.00-20 -15 -10 -5 0 5 10 15 20Reference Pressure (kPa)The standard pressure equation is a quadratic polynomial in frequency squared. We noteda somewhat systematic deviation between the test unit and the reference pressure thatlooks like an S-curve. We added a small cubic term to the pressure equation to bestmodel the sensor. The residual rms error was 0.2 Pa, including repeatability andhysteresis. The shown data is the same as in the previous plot.Best Conformance Fit to CEC Standard202Bg Deviation (Pa)3.002.251.500.750.00-0.75-1.50-2.25-3.00-20 -15 -10 -5 0 5 10 15 20Reference Pressure (kPa)