12.07.2015 Views

Low Level Measurements Handbook

Low Level Measurements Handbook

Low Level Measurements Handbook

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

FIGURE 1-2: Theoretical Limits of Voltage <strong>Measurements</strong>1kV10 3NoiseVoltage1VWithin theoretical limits10 01mV1µV1nV1pVNear theoretical limitsProhibitedby noise10 –310 –610 –910 –1210 0 10 3 10 6 10 9 10 121Ω 1kΩ 1MΩ 1GΩ 1TΩSource Resistanceinput offset current 1 when measuring voltage and lower input resistancecompared to more sensitive instruments intended for low level DC measurements.These characteristics cause errors in the measurement; refer toSections 2 and 3 for further discussion of them.Given these DMM characteristics, it’s not possible to use a DMM tomeasure signals at levels close to theoretical measurement limits, as shownin Figure 1-3. However, if the source resistance is 1MΩ or less, or if thedesired resolution is no better than 0.1µV (with low source resistance), thesignal level isn’t “near theoretical limits,” and a DMM is adequate. If bettervoltage sensitivity is desired, and the source resistance is low (as it must bebecause of theoretical limitations), a nanovoltmeter provides a means ofmeasuring at levels much closer to the theoretical limits of measurement.With very high source resistance values (for example, 1TΩ), a DMM isn’t asuitable voltmeter. DMM input resistance ranges from 10MΩ to 10GΩ—severalorders of magnitude less than a 1TΩ source resistance, resulting insevere input loading errors. Also, input currents are typically manypicoamps, creating large voltage offsets. However, because of its much higherinput resistance, an electrometer can make voltage measurements at levelsthat approach theoretical limits. A similar situation exists for low levelcurrent measurements; DMMs generally have a high input voltage drop1Input current flows in the input lead of an active device or instrument. With voltage measurements, theinput current is ideally zero; thus, any input current represents an error. With current measurements, thesignal current becomes the input current of the measuring instrument. However, some background currentis always present when no signal current is applied to the instrument input. This unwanted currentis the input offset current (often called just the offset current) of the instrument.The source and test connections can also generate unwanted offset currents and offset voltages.A leakage current is another unwanted error current resulting from voltage across an undesired resistancepath (called leakage resistance). This current, combined with the offset current, is the total errorcurrent.1-4 SECTION 1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!