FIGURE 1-27: DMM Block DiagramACAttenuatorACConverterDigitalDisplayHIINPUTAmpsACDCOhmsDCAttenuatorOhmsConverterACDCOhmsA/DConverterPrecisionReferenceDigitalOutputPorts(IEEE-488,RS-232,Ethernet)PrecisionShuntsLOverter provides a DC analog signal for resistance measurements. Precisionshunts are used to convert currents to voltages for the amps functions.Once the input signal is appropriately processed, it’s converted to digitalinformation by the A/D converter. Digital data is then sent to the displayand to the digital output port (IEEE-488, RS-232, or Ethernet).NanovoltmetersA nanovoltmeter is a sensitive voltmeter optimized to measure very low voltages.As shown in Figure 1-28, the nanovoltmeter incorporates a low noisepreamplifier, which amplifies the signal to a level suitable for A/D conversion(typically 2–3V full scale). Specially designed preamplifier circuitsensure that unwanted noise, thermoelectric EMFs, and offsets are kept to anabsolute minimum.FIGURE 1-28: Typical NanovoltmeterHIDCV InputLO<strong>Low</strong>-NoisePreamplifierRangeSwitchingA/DConverterDisplayIEEE-488,RS-232OffsetCompensationMicroprocessor1-30 SECTION 1