12.07.2015 Views

Thermo Scientific Hypercarb Columns

Thermo Scientific Hypercarb Columns

Thermo Scientific Hypercarb Columns

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

F OOD AND B EVERAGEFigure 4: Column coating lifetime. Reproducible retention times for 16000 column volumes of mobile phase through the column or 445 injections.Column: <strong>Hypercarb</strong> 5 µm, 100 x 4.6 mm (CTA 8 mM); Mobile phase: 2 mM Na 2 CO 3 /1 mM NaHCO 3 + 5% MeCN; other conditions as described in the experimental section.ConclusionDynamically CTA coated <strong>Hypercarb</strong> columns possess goodstability, reproducibility, linearity and lifetime performanceworking with organic compositions up to 25%. This approachoffers an alternative to the conventional methodology forthe determination on inorganic anions in drinking water.Application NotesReferences1. C. Elfakir, P. Chaimbault, M. Dreux, J. Chromatogr. A, 829 (1998) 1932. K. Petritis, P. Chaimbault, C. Elfakir, M. Dreux, J. Chromatogr. A, 1999,833, 1473. T. Takeuchi, T. Kojima, T. Miwa, J. High, Resol. Chromatogr., 2000, 23, 5904. T. Okamoto, A. Isozaki, H. Nagashima, J. Chromatogr. A, 1998, 800, 2395. H. Nagashima, T. Okamoto, J. Chromatogr. A, 1999, 855, 261-266Figure 5: Analysis of bottled drinking water.39