Transducer Electrical CharacteristicsThe basic Digiquartz® transducers, Series, 2000, 3000, 4000, and5300, provide frequency output signals for pressure and temperature.The nominal pressure signal frequency is a 10% change within theband of 30 kHz to 42 kHz. The temperature signal is a nominal 45ppm/°C sensitivity within the band of 168 kHz to 172 kHz. Theyoperate from 6 to 25 VDC at 1.3 mA and are reverse polarityprotected.Current drain increases when driving long lines or low impedances.Refer to the SCD supplied with the transducer for any special inputpower options or considerations.Standard wire color code for a basic Digiquartz® transducer is asfollows:REDBLACKBLUEWHITE6 to 25 VDCSignal and power groundPressure signal, 30 to 42 kHzTemperature signal, 168 to 172 kHzThe pressure and temperature signal wires are configured as twistedpairs. Each signal line is twisted with a negative (black) wire tominimize cross coupling between the temperature and pressuresignals. If the signal lines are to be extended, the black wires shouldbe continued and twisted with the signal wire at a rate of 4 to 5 turnsper inch.4