TABLE 8GC OPERATING CONDITIONS FOR ORGANOCHLORINE PESTICIDESFOR THE DUAL COLUMN METHOD OF ANALYSISHIGH TEMPERATURE, THICK FILMColumn 1:Column 2:Carrier gas:Carrier gas flowrate:Makeup gas:Makeup gas flowrate:Injector temperature:Detector temperature:Initial temperature:Temperature program:Final temperatureDB-1701 (J&W) or equivalent30 m x 0.53 mm ID1.0 µmDB-5 (J&W) or equivalent30 m x 0.53 mm ID1.5 µmHelium6 mL/minuteNitrogen20 (mL/min)250EC320EC150EC, hold 0.5 min150EC to 190EC at 12EC/min, hold 2 min190EC to 275EC at 4EC/min275EC, hold 10 minCD-ROM <strong>8081A</strong> - 33 Revision 1December 1996