Command strings Value of Argument () Action# 0 [:][;] Software Reset# 1:[;] 1...4095 Travel lens2 to absolute position# 2:[;] 1...4095 Travel lens1 to absolute position# 3:[;] -9...0...9 Travel lens2 with correspondingspeed and turn direction (+/-) untilstop command(0 = No travel; 9 = Max. speed)# 5:[;] -9...0...9 Travel lens1 with correspondingspeed and turn direction (+/-) untilstop command(0 = No travel; 9 = Max. speed)# 7 [:][;] Position query lens2;Response: 1...4095# 8 [:][;] Position query lens1;Response: 1...4095# 9:;[;] 1...4095 in each case Travel lens1 to ‘Value 1’ andlens2 to ‘Value 2’# 12:[;] address 1...512 read stored value fromcontroller# 13:[;] address 1...512 read stored value from BeamExpander# 22:[:][;] integer value / 100 read offset value lens1# 23:[:][;] integer value / 100 read offset value lens2# 50:[;] '0' is off, '1' is on serial echo on / off# 50:[;] query serial echo on / off# 51:[;] '0' is off, '1' is on advanced mode on / off# 51:[;] query advanced mode on / off# 54:[;] 1200 2400 4800 9600 1920038400adjust communicationbaudrate# 253:[:][;] Output Copyright and Version# 254:[:][;] Stop commandCharacters in '[]'brackets may be omitted.All not mentioned command codes and sequences are reserved and should be avoided.27