Error messages from theBeam Expander ControllerE00E01E02E03E04E05E06E10E11E50MExMeaninginvalid commandopen connection to position sensing lens1open connection to position sensing lens2shorted connection to position sensing lens1shorted connection to position sensing lens2measuring system lens1 defectivemeasuring system lens2 defectiveno movement of lens1 detectedno movement of lens2 detectedserial buffer overrunchecksum error in Beam Expander EEPROMAll not mentioned error codes are reserved.The non-volatile memory of the Beam ExpanderInside the Beam Expander, is a non volatile memory used for storing necessary values for thecontrolling of the Beam Expander. In case of building a proprietary control electronics, it is necessary toknow how to access the stored data. Detailed information on how to access these data is available onrequest from LINOS Photonics GmbH & Co. KG.29